Trilogy200 clinical manual
108
Alarm Priority Audible Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
Ventilator
Service
Required
High • • • • • Red ashing button;
“Ventilator Service
Required” message
Operates User: Press Alarm Indicator/
Audio Pause button to silence
alarm. Or, press Left (Reset)
button to reset alarm. The
alarm goes away until next
reminder appears. Connect
patient to alternate source of
ventilation and contact your
health care professional as
soon as possible.
Clinician: If alarm continues,
have the device serviced.
Check Circuit High • • • • • Red ashing button;
“Check Circuit”
message
Operates User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press
Left (Reset) button to reset
alarm. Check patient circuit
to make sure it is set up as
your health care professional
specied. Make sure there
is no kinked or pinched
tubing. Make sure tubing is
properly attached. Make sure
all tubing and exhalation
devices are clean and free
of condensation (refer to
the Troubleshooting section
– tubing condensation). If
using Active PAP or Active
Flow Exhalation Device, make
sure diaphragm is inserted
correctly and is not wrinkled
or kinked. If alarm continues,
contact your health care
professional.
Clinician: Verify patient
status. Verify that breathing
circuit setup is correct.
Correct any errors. If alarm
continues, have device
serviced.