65

If Detailed View is turned on in the Options menu, the Monitor screen will look like the screen shown below.

Monitor Panel

Measured Settings

Panel

Status Panel

This screen contains more detailed information about the therapy.

Monitor Screen – Detailed View On

The top Monitor panel contains the Prescription indicator if a dual prescription exists, the therapy mode, a graph displaying the current pressure, and the current date and time. Additionally, this panel also displays patient pressure, respiratory rate (RR), exhaled tidal volume (Vte), and leak.

The second panel in Detailed view is the Measured Settings panel. It provides patient-related data including Peak Inspiratory Pressure (PIP), Minute Ventilation, Peak Inspiratory Flow, Mean Airway Pressure (MAP), and I:E Ratio.

The third panel is the Status panel and shows the same information displayed in the Detailed View Off screen, including features in use such as Ramp and battery status.

Chapter 5 Viewing and Changing Settings

Page 76
Image 76
Philips Trilogy 200 manual Monitor Screen Detailed View On