Trilogy200 clinical manual
146
Setting Value
PEEP 4 cm H2O
Sigh O
Circuit Disconnect O
Low Inspiratory Pressure 6 cm H2O
High Inspiratory Pressure 10 cm H2O
Apnea O
All other alarms O
B. Verify the Alarm
Wait up to 40 seconds and verify the following alarm signals:
The Medium Priority audible indicator sounds A yellow light ashes on the Alarm Indicator/Audio
Pause button
The High Inspiratory Pressure alarm condition
appears on the screen, highlighted in yellow
C. Modify Ventilator Alarm Settings
Modify the High Inspiratory Pressure setting to match the one
shown below in Table 12.
Setting Value
High Inspiratory Pressure 60 cm H2O
D. Verify Reset
Wait 40 seconds and verify the following auto-reset conditions:
The Medium Priority audible indicator has stopped
sounding
The yellow light on the Alarm Indicator/Audio Pause
button has stopped ashing
6. Verify the Low Inspiratory Pressure Alarm
This procedure veries that the Low Inspiratory Pressure alarm
is working properly. It assumes that you have attached the test
lung, veried ventilator settings, and turned on ventilator power as
described in the Initial Setup section.
Table 12 - Modify High
Inspiratory Pressure Alarm
Setting