Control Accuracy

Parameter

Range

 

 

Accuracy

 

 

IPAP

4 to 501 cm H2O2

Greater of 2 cm H2O or 8% of setting

 

 

 

EPAP

0 to 25 cm H2O for Active Circuits

Greater of 2 cm H2O or 8% of setting

 

4 to 25 cm H2O for Passive Circuits

 

 

 

CPAP

4 to 20 cm H2O

Greater of 2 cm H2O or 8% of setting

PEEP

0 to 25 cm H

2

O for Active Circuits

Greater of 2 cm H

2

O or 8% of setting4

 

4 to 25 cm H2O for Passive Circuits

 

 

 

Pressure

4 to 50 cm H2O

Greater of 2 cm H2O or 8% of setting

Pressure Support

0 to 30 cm H

2

O3

Greater of 2 cm H

2

O or 8% of setting4

Tidal Volume

50 to 2000 ml5

Greater of 15ml or 10% of setting (Active

 

 

 

 

Circuits)

 

 

 

 

 

 

Greater of 15ml or 15% of setting (Passive

 

 

 

 

Circuits)

 

 

Breath Rate

0 to 60 BPM for AC mode

Greater of ± 1 BPM or ±10% of the setting

 

1 to 60 BPM for all other modes

 

 

 

Timed Inspiration

0.3 to 5.0 seconds

± .1 second

 

 

 

 

 

 

 

Rise Time

1 to 66

 

 

±2 cm H2O (the device will increase to a

 

 

 

 

pressure of .67* (IPAP - EPAP) ±2 cm H2O @ the

 

 

 

 

set rise time multiplied by 100 ms for pressure

 

 

 

 

supports less than or equal to 25.)

Ramp Start Pressure

0 to 25 cm H2O for Active Circuits

8% of setting + 2% Full Scale

 

4 to 25 cm H2O for Passive Circuits

 

 

 

 

4 to 19 cm H2O in CPAP mode

 

 

 

Ramp Length

Off, 5 to 45 minutes

±2 minutes

 

 

 

 

 

 

 

 

 

Flex

Off, 1 to 37

 

 

N/A

 

 

Flow Trigger Sensitivity

1 to 9 l/min

 

 

N/A

 

 

 

 

 

 

 

 

 

Flow Cycle

10 to 90%

 

 

N/A

 

 

 

 

 

 

 

Apnea Rate

4 to 60 BPM

 

 

Greater of 1 BPM or 10% of setting

 

 

 

 

 

 

 

Specifications listed are based on using a standard circuit consisting of 1.8 meter tubing- REF 622038, one of the following exhalation devices: Whisper Swivel II - REF 332113; Active PAP Exhalation Device - REF 1053716; Active Flow Exhalation Device - REF 1049503 or Active Exhalation Device – REF 1065658; UPB Active PAP Exhalation Device - REF 1073862 and

Trilogy Proximal Flow Sensor- REF 1050408 (if required).

1Limited to 25 cm H2O when using the Bi-Flex feature in S mode.

2Pressure units may be cm H2O, hPa, or mBar depending on device setup.

3The difference between the Inspiratory Pressure and the Expiratory Pressure must never be more than 30 cm H2O.

4Pressure Support and PEEP not to exceed 50 cm H2O.

5All flows and volumes are measured at BTPS conditions.

6The range of values correspond to tenths of seconds (e.g., a setting of 4 indicates a Rise Time of 0.4 seconds).

7Flex is not available when AVAPS is active. Flex is not available with Active Circuits.