112

Alarm

Priority

Audible

Visual (Alarm

Device

User/Clinician Action

 

 

 

Indicator Button

Action

 

 

 

 

and Display)

 

 

High

High

• • •

• •

Red flashing button;

Operates

User: Press Alarm Indicator/

 

 

Respiratory

 

 

 

“High Respiratory

 

Audio Pause button to

Rate

 

 

 

Rate” message

 

silence alarm. Or, press Left

 

 

 

 

 

 

(Reset) button to reset alarm.

 

 

 

 

 

 

Continue using device. If

 

 

 

 

 

 

alarm continues, contact your

 

 

 

 

 

 

health care professional.

 

 

 

 

 

 

Clinician: Verify patient

 

 

 

 

 

 

status.

 

 

 

 

 

 

 

Low Respiratory

High

• • •

• •

Red flashing button;

Operates

User: Press Alarm Indicator/

 

 

Rate

 

 

 

“Low Respiratory

 

Audio Pause button to

 

 

 

 

Rate” message

 

silence alarm. Or, press Left

 

 

 

 

 

 

(Reset) button to reset alarm.

 

 

 

 

 

 

Continue using device. If

 

 

 

 

 

 

alarm continues, contact your

 

 

 

 

 

 

health care professional.

 

 

 

 

 

 

Clinician: Verify patient

 

 

 

 

 

 

status.

 

 

 

 

 

 

 

High

Escalates

(for first two

Peak Pressure symbol

Operates

User: Press Alarm Indicator/

Inspiratory

from

turns red. When

 

Audio Pause button to silence

consecutive

 

Pressure

Audible

condition first occurs,

 

alarm. Or, press Left (Reset)

occurrences)

 

 

indicator

a beep will sound.

 

 

 

• • •

(for 3rd

 

button to reset alarm. Make

 

to Medium

When condition

 

 

consecutive

 

sure tubing is not pinched or

 

priority and

occurs for third time,

 

 

occurrence)

 

blocked. If alarm continues,

 

then High

button flashes yellow

 

 

• • •

• •

 

contact your health care

 

priority

and yellow “High

 

 

(for 10th

 

professional.

 

 

Inspiratory Pressure”

 

 

 

consecutive

 

 

 

message appears.

 

 

 

 

occurrence)

 

Clinician: Verify patient

 

 

When condition

 

 

 

 

 

 

 

 

 

 

occurs for 10th

 

status. If problem continues,

 

 

 

 

time, button flashes

 

have device serviced.

 

 

 

 

red and red “High

 

 

 

 

 

 

Inspiratory Pressure”

 

 

 

 

 

 

message appears.

 

 

Trilogy200 clinical manual

Page 123
Image 123
Philips Trilogy 200 manual 112