142

2. Verify the High Tidal Volume Alarm

This procedure verifies that the High Tidal Volume alarm is working properly. For Passive and Active Flow circuits, this will verify the High Vte alarm. For Active PAP circuits, this will verify the High Vti alarm. It assumes that you have attached the test lung, verified the ventilator settings, and turned on ventilator power as described in the Initial Setup section.

A. Change Alarm Ventilator Setting

Modify the High Tidal Volume alarm setting to match the one shown below in Table 3.

Setting

Value

High Vte/High Vti

50 ml

B.Verify the Alarm

Wait up to 40 seconds and verify the following alarm signals:

The High Priority audible indicator sounds

A red light flashes on the Alarm Indicator/Audio Pause button

The High Tidal Volume alarm condition appears on the screen, highlighted in red

C.Modify Ventilator Alarm Settings

Modify the High Tidal Volume alarm setting to match the one shown below in Table 4.

Table 3 - High Vte/Vti Alarm Setting

Note: Do not use the “Reset” button to manually reset the alarm. Instead, use the “Modify” button to change ventilator settings. This applies to all tests.

Setting

Value

High Vte/High Vti

500 ml

 

 

D.Verify Reset

Wait 40 seconds and verify the following auto-reset conditions:

The High Priority audible indicator has stopped sounding

The red light on the Alarm Indicator/Audio Pause button has stopped flashing

Table 4 - Modify Vte/High Vti Alarm Setting

Trilogy200 clinical manual

Page 153
Image 153