ResMed III ST-A user manual Testing the Alarm, Replacing the Air Filter, Servicing

Page 26

CAUTION

! Do not attempt to open the VPAP III ST-A. There are no user serviceable parts inside. Repairs and internal servicing should only be performed by an authorized service agent.

Testing the Alarm

Test the alarm weekly. See “Testing the Alarm” on page 29.

Replacing the Air Filter

Please refer to the illustrations in section I of the illustration sheet.

Inspect the air filter every month to check if it is blocked by dirt or contains holes. With normal use of a VPAP III ST-A, the air filter needs to be replaced every six months (or more often if your device is in a dusty environment). To replace the air filter:

1.Remove the air filter cover at the back of the VPAP III ST-A (I-1).

2.Remove and discard the old air filter.

3.Insert a new filter with the blue tinted side facing out (I-2).

4.Replace the air filter cover (I-3).

WARNING

! Do not wash the air filter. The air filter is not washable or reusable.

Note: The air filter should be inspected once a month.

Servicing

This product (VPAP III ST-A) should be inspected by an authorized ResMed service centre 5 years from the date of manufacture. Prior to this, the device is intended to provide safe and reliable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed. Applicable ResMed warranty details are provided with the device at the time of original supply. Of course, as with all electrical devices, if any irregularity becomes apparent, you should exercise caution and have the device inspected by an authorized ResMed service centre.

If you feel that your VPAP III ST-A is not performing properly, see “Troubleshooting” on page 35.

CAUTION

! Inspection and repair should only be performed by an authorized agent. Under no circumstances should you attempt to service or repair the flow generator yourself.

18

Image 26
Contents Vpap III ST-A ROW ENG Illustrations System Components52 cm 3 m 248139/3 05 Manufactured by Contents System Specifications Limited Warranty Index User/Owner Responsibility IntroductionMedical Information What the Vpap III ST-A is Intended forPage Adverse Effects Page Vpap III ST-A System MasksAccessories HumidifiersSetting up the Vpap III ST-A System Setting up the Vpap III ST-AAttaching a HumidAire 2iC Humidifier Attaching a HumidifierAttaching a HumidAire 2i Humidifier Attaching a HumidAire HumidifierAttaching a Passover Humidifier LEDs Using the LCD Screen and KeypadLCD Screen Keypad Keys Vpap III ST-A keypad has the following keysPage Starting Treatment Ramp TimeStopping Treatment Using the HumidAire 2i Warm-up FeatureUsing the Mask-Fitting Feature High Leak Page Weekly Cleaning Cleaning and MaintenanceDaily Cleaning Periodic CleaningServicing Testing the AlarmReplacing the Air Filter Ramp Screen How to Use the Vpap III ST-A MenusHow to Use the Standard Menu How to Use the Detailed Menu if enabled by your clinician Settings Summary ScreensMIRAGE, MIR Full STANDARD, Ultra Settings Menu Menu Functions Detailed Menu onlySmartStart Results Menu Options MenuTreatment Screens Servicing MenuOxygen saturation This screen only Level SpO2 98% Non-vented Mask Alarm Screen Page Alarm Mute Key AlarmsAlarms Troubleshooting Testing the AlarmLCD Ipap Lower LCD System Error LCD LOW MV Alarm Starting out Using a Battery to Power the Vpap III ST-AHelpful Hints Nasal IrritationResMed DC-24/30 Converter Troubleshooting No displaySystem Error System Specifications Dynamic pressure characteristicsIEC 60601-1 Classifications Pressure Volume Curve Pressure VariationSymbols which may Appear on the Product Environmental informationEmissions test Compliance Next Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Page Index Index

III ST-A specifications

The ResMed III ST-A is a state-of-the-art ventilator specifically designed for patients with respiratory challenges, particularly those suffering from obstructive sleep apnea, neuromuscular conditions, or chronic obstructive pulmonary disease (COPD). It stands out due to its user-friendly features, advanced technologies, and robust performance, making it a preferred choice for both clinicians and patients.

One of the main features of the ResMed III ST-A is its intelligent algorithm designed to optimize minute ventilation. This provides variable support that adapts to the patient's breathing patterns, enhancing comfort and ensuring effective ventilation. By offering a range of modes, including spontaneous timed (ST) and volume-assured pressure support (VAPS), this device caters to various clinical needs, facilitating customized therapy based on individual patient requirements.

The built-in leak compensation feature of the ResMed III ST-A allows it to maintain accurate pressure delivery even in the presence of large leaks, which is crucial for ensuring effective therapy. This is particularly beneficial for patients who may have difficulty achieving a proper seal with their masks.

Incorporating advanced technologies, the ResMed III ST-A features AutoSet technology, which automatically adjusts pressure levels in response to detected changes in airflow. This promotes optimal treatment at all times, reducing the burden on patients and allowing for a more seamless sleep experience. The device also supports a range of connectivity options, enabling healthcare providers to monitor patient compliance and therapy effectiveness remotely through ResMed's cloud-based platform.

The ResMed III ST-A is designed with a user-friendly interface, including a bright, easy-to-read display and simplified controls, promoting ease of operation for both healthcare professionals and patients. Its compact size and lightweight design enhance portability, making it suitable for home and travel use.

Furthermore, the device's comprehensive data logging capabilities allow clinicians to access a wealth of information regarding patient adherence and sleep patterns, thus supporting informed decision-making. The ResMed III ST-A also prioritizes patient safety, featuring alarms and alerts for high- and low-pressure conditions.

In conclusion, the ResMed III ST-A embodies cutting-edge technology and thoughtful design, providing versatile and effective respiratory support. With its combination of intelligent algorithms, user-friendly features, and robust patient monitoring capabilities, it is well-positioned to enhance the quality of life for individuals with respiratory challenges.