ResMed III ST-A user manual Alarms Troubleshooting, Alarm Mute Key, Testing the Alarm

Page 37

The Alarms

The VPAP III ST-A unit is fitted with alarms to alert you to changes that will affect your treatment.

Alarm Mute Key

You can mute an alarm by pressing the Alarm Mute key once. If the problem is still present, the alarm will sound again after two minutes. An alarm LED will remain lit for as long as the problem is present.

Testing the Alarm

To test the alarm, hold down the Alarm Mute key for three seconds. On release of the Alarm Mute key, the alarm should sound and the LED should flash for 10 seconds and then stop. ResMed recommends testing the alarm each week.

All menus remain available during an alarm. Treatment screens are not viewable during an alarm condition.

LCD screen

Alarm LEDs

Alarm Mute key

Alarms Troubleshooting

The most common reason for an alarm to sound is because the system has not been properly assembled. Check that the air tubing has been properly attached to the flow generator and mask (and humidifier if used).

When an alarm has been activated, the bottom line of the LCD screen will display either alarm information or recommended instructions for you or your clinician.

Note: The alarm actions listed below are based on having the appropriate alarm settings for your therapy. When an alarm is activated, confirm the alarm settings with your clinician.

CAUTION

! In the event of power failure or machine malfunction, remove the mask.

The Alarms

29

Image 37
Contents Vpap III ST-A ROW ENG System Components Illustrations52 cm 3 m 248139/3 05 Manufactured by Contents System Specifications Limited Warranty Index Medical Information IntroductionUser/Owner Responsibility What the Vpap III ST-A is Intended forPage Adverse Effects Page Masks Vpap III ST-A SystemHumidifiers AccessoriesSetting up the Vpap III ST-A Setting up the Vpap III ST-A SystemAttaching a HumidAire 2i Humidifier Attaching a HumidifierAttaching a HumidAire 2iC Humidifier Attaching a HumidAire HumidifierAttaching a Passover Humidifier LCD Screen Using the LCD Screen and KeypadLEDs Vpap III ST-A keypad has the following keys Keypad KeysPage Ramp Time Starting TreatmentUsing the Mask-Fitting Feature Using the HumidAire 2i Warm-up FeatureStopping Treatment High Leak Page Daily Cleaning Cleaning and MaintenanceWeekly Cleaning Periodic CleaningReplacing the Air Filter Testing the AlarmServicing How to Use the Standard Menu How to Use the Vpap III ST-A MenusRamp Screen Settings Summary Screens How to Use the Detailed Menu if enabled by your clinicianMIRAGE, MIR Full STANDARD, Ultra Menu Functions Detailed Menu only Settings MenuSmartStart Options Menu Results MenuServicing Menu Treatment ScreensOxygen saturation This screen only Level SpO2 98% Non-vented Mask Alarm Screen Page Alarms Troubleshooting AlarmsAlarm Mute Key Testing the AlarmLCD Ipap Lower LCD System Error LCD LOW MV Alarm Helpful Hints Using a Battery to Power the Vpap III ST-AStarting out Nasal IrritationResMed DC-24/30 Converter No display TroubleshootingSystem Error Dynamic pressure characteristics System SpecificationsIEC 60601-1 Classifications Pressure Variation Pressure Volume CurveEnvironmental information Symbols which may Appear on the ProductEmissions test Compliance Next Portable and mobile RF communications Separation distance according to frequency of transmitter Limited Warranty Page Index Index

III ST-A specifications

The ResMed III ST-A is a state-of-the-art ventilator specifically designed for patients with respiratory challenges, particularly those suffering from obstructive sleep apnea, neuromuscular conditions, or chronic obstructive pulmonary disease (COPD). It stands out due to its user-friendly features, advanced technologies, and robust performance, making it a preferred choice for both clinicians and patients.

One of the main features of the ResMed III ST-A is its intelligent algorithm designed to optimize minute ventilation. This provides variable support that adapts to the patient's breathing patterns, enhancing comfort and ensuring effective ventilation. By offering a range of modes, including spontaneous timed (ST) and volume-assured pressure support (VAPS), this device caters to various clinical needs, facilitating customized therapy based on individual patient requirements.

The built-in leak compensation feature of the ResMed III ST-A allows it to maintain accurate pressure delivery even in the presence of large leaks, which is crucial for ensuring effective therapy. This is particularly beneficial for patients who may have difficulty achieving a proper seal with their masks.

Incorporating advanced technologies, the ResMed III ST-A features AutoSet technology, which automatically adjusts pressure levels in response to detected changes in airflow. This promotes optimal treatment at all times, reducing the burden on patients and allowing for a more seamless sleep experience. The device also supports a range of connectivity options, enabling healthcare providers to monitor patient compliance and therapy effectiveness remotely through ResMed's cloud-based platform.

The ResMed III ST-A is designed with a user-friendly interface, including a bright, easy-to-read display and simplified controls, promoting ease of operation for both healthcare professionals and patients. Its compact size and lightweight design enhance portability, making it suitable for home and travel use.

Furthermore, the device's comprehensive data logging capabilities allow clinicians to access a wealth of information regarding patient adherence and sleep patterns, thus supporting informed decision-making. The ResMed III ST-A also prioritizes patient safety, featuring alarms and alerts for high- and low-pressure conditions.

In conclusion, the ResMed III ST-A embodies cutting-edge technology and thoughtful design, providing versatile and effective respiratory support. With its combination of intelligent algorithms, user-friendly features, and robust patient monitoring capabilities, it is well-positioned to enhance the quality of life for individuals with respiratory challenges.