5.0SERVICE

5.1Sensor Replacement

The gas sensor needs replacement when:

-It is no longer possible to obtain correct calibration

-The failed sensor alarm (oscillating red/green LED) is on

-The sensor output signal is noisy, causing incorrect gas alarms.

To replace the sensor:

1.Confirm that the system power has been removed.

2.Remove the cover from the main enclosure.

3.Unplug the sensor connector on the lower right hand corner of the electronics board.

4.Unscrew the sensor assembly from the end of the enclosure.

5.Reverse the preceding steps to install the new sen- sorassembly.

6.Restore power and allow a minimum of 24 hours for stabilization before recalibration. The calibra- tion should be re-verified after 6 days of continu- ous operation.

Figure 5-1 Model 2050/2060 Electronics Board Connections

Page: 7

2050/2060 Gas Monitors (5/04)

Image 10
Contents GAS Monitors Applicability & Effectivity Applicability & EffectivityService Table of Contents Section Product DescriptionAlarm Calibration InstallationModel Gas Monitor ConfigurationsIntroduction TTLTemperature Range Overview WiringConfiguration Model 2050-2060 SeriesBuzzer Relay Fail Safe Closed Open Trouble Alarm Warm-Up AlarmBuzzer Relay Fail Safe On/Off Oscilating Gas Alarm Buzzer Relay Fail Safe On/Off Oscilating Outputs3Alarm Calibration Check Process Factory Alarm CalibrationFrequency of Alarm Calibration Check Model Time Gas Cc/min4EquipmentRequired Alarm Calibration ProcedureBe sure that the area is non-hazardous be- foreproceeding Sensor Replacement Model 2050/2060 Electronics Board ConnectionsGas Sensor Module Locations Alarm DevicesMounting Interconnect Wiring Table 2050/2060 Gas Monitors 5/04 Alarm Configuration Relay OperationAlarm Panels Replacement parts Calibration Accessories