Sierra Monitor Corporation 2060-10, 2060-00, 2050-10, 2050-00 Alarm Configuration, Relay Operation

Page 13

Instruction Manual

7.0 ALARM CONFIGURATION

Model 20X or 200X Configured w ith Alarm Relay Output

Configuration Jum pers

Buzzer Active

Relay

Operation

 

 

Y es

No

Normally

 

Normally

Not Energized

 

Energized

 

 

 

Default

 

Default

 

 

Install Jumper

Remove Jumper

Install Jumper

 

Install Jumper

JP6

JP6

JP2 pins 1-2

 

JP2 pins 2-3

 

(Best for 20X)

 

 

 

 

 

 

 

 

 

 

 

 

 

W iring Terminations

 

Power Supply

 

 

Output Terminals

 

TB1-1

TB1-2

TB2-1

 

TB2-2

 

TB2-3

+ DCV

0 DC

NC

 

Common

 

NO

(9-24 VDC)

Common

 

 

 

 

 

Model 20X or 200X Configured for Interface to Model 2102 Alarm Panel

 

 

Configuration Jum pers

 

Buzzer Active

Relay

Operation

 

Interface Selection

Yes

No

Normally

 

Normally

 

Model 2102

Not Energized

 

Energized

 

 

 

 

 

 

Default

 

Default

 

 

 

 

Install Jumper

Remove Jumper

N/A

 

N/A

 

Install Jumpers

JP6

JP6

 

 

 

 

JP3 pins 1-2

 

(Best for 20X)

 

 

 

 

JP2 pins 1-2

 

 

 

 

 

 

 

 

W iring Terminations

 

Power Supply

 

 

Output Terminals

 

TB1-1

TB1-2

TB2-1

 

TB2-3

 

TB2-2

+ DCV

0 DC

Safe

 

Alarm

 

 

To Model 2102

To Model 2102

To Model 2102

 

To Model 2102

 

Not Used

Terminal J5 -2

Terminal J5 -3

Terminal J5 -1

 

Terminal J5 -4 or

 

or J6-2

or J6-3

or J6-1

 

J6-4

 

 

 

 

 

 

 

 

 

Table 6.2

Jumper Configuration and W iring Terminations

Relay Operation

Normally Not Energized

 

TB2-1 NC

TB2-2 Common

TB2-3 NO

No Power

Connected

Open

Sensor Fail

 

Oscillating

 

Safe

Connected

Open

Alarm

Open

Connected

 

Normally Energized - Fail Safe

 

 

TB2-1 NC

TB2-2 Common

TB2-3 NO

No Power

Connected

Open

Sensor Fail

 

Oscillating

 

Safe

Open

Connected

Alarm

Connected

Open

Table 6.3

Relay Operating States

2050/2060 Gas Monitors (5/04)

Page: 10

Image 13
Contents GAS Monitors Applicability & Effectivity Applicability & EffectivityAlarm Calibration Table of Contents Section Product DescriptionService InstallationIntroduction Gas Monitor ConfigurationsModel TTLTemperature Range Configuration WiringOverview Model 2050-2060 SeriesBuzzer Relay Fail Safe On/Off Oscilating Gas Alarm Warm-Up AlarmBuzzer Relay Fail Safe Closed Open Trouble Alarm Buzzer Relay Fail Safe On/Off Oscilating OutputsFrequency of Alarm Calibration Check Factory Alarm Calibration3Alarm Calibration Check Process Model Time Gas Cc/min4EquipmentRequired Alarm Calibration ProcedureBe sure that the area is non-hazardous be- foreproceeding Model 2050/2060 Electronics Board Connections Sensor ReplacementGas Sensor Module Locations Alarm DevicesMounting Interconnect Wiring Table 2050/2060 Gas Monitors 5/04 Relay Operation Alarm ConfigurationCalibration Accessories Alarm Panels Replacement parts