Sierra Monitor Corporation 2060-10, 2060-00 manual 4EquipmentRequired, Alarm Calibration Procedure

Page 9

Instruction Manual

4.4EquipmentRequired

The following tools and equipment will be required for calibration:

-Jewelers Screwdriver

-Calibration Gas

-Calibration gas delivery system

For accurate calibration use a gas mixture at the re- quired concentration mixed in an air balance, rather than with an inert gas like nitrogen. This gas and the required delivery equipment such as the Model 1200- 26 Calibrator is available from Sierra Monitor Corpo- ration.

4.5Alarm Calibration Procedure

1.The monitor should be in the safe condition prior to calibration (green LED “ON”).

-Be sure that the area is non-hazardous be- foreproceeding.

2.With the monitor in operation, expose the sensor to a sample of calibration gas. Apply calibration gas directly to the sensor insuring that the delivery cup loosely covers the sensor. Use 80 PPM H2 for the Model 2050 and 150 PPM CO for the Model 2060.

3.The red LED on the face of the monitor should turn "on". If the alarm occurs within 18-20 seconds and stops within one minute of the removal of the cali- bration gas, the monitor is in calibration and re- quires no further adjustment.

4.If the monitor fails to alarm within 90 seconds of the application of the gas adjust the potentiometer counter clockwise until the alarm turns "on".

5.If the monitor fails to stop alarming within one minute of the removal of the gas adjust the poten- tiometer clockwise until the alarm stops.

6.After adjustment, repeat steps 4 and 5 to verify cali- bration.

7.When the calibration is complete, reconnect any alarm equipment as necessary.

2050/2060 Gas Monitors (5/04)

Page: 6

Image 9
Contents GAS Monitors Applicability & Effectivity Applicability & EffectivityAlarm Calibration Table of Contents Section Product DescriptionService InstallationIntroduction Gas Monitor ConfigurationsModel TTLTemperature Range Configuration WiringOverview Model 2050-2060 SeriesBuzzer Relay Fail Safe On/Off Oscilating Gas Alarm Warm-Up AlarmBuzzer Relay Fail Safe Closed Open Trouble Alarm Buzzer Relay Fail Safe On/Off Oscilating OutputsFrequency of Alarm Calibration Check Factory Alarm Calibration3Alarm Calibration Check Process Model Time Gas Cc/minAlarm Calibration Procedure 4EquipmentRequiredBe sure that the area is non-hazardous be- foreproceeding Model 2050/2060 Electronics Board Connections Sensor ReplacementAlarm Devices Gas Sensor Module LocationsMounting Interconnect Wiring Table 2050/2060 Gas Monitors 5/04 Relay Operation Alarm ConfigurationCalibration Accessories Alarm Panels Replacement parts