Bacharach 3015-4256 manual Electrical Wiring

Page 13

Electrical Wiring

The HGMSZ uses a universal power supply that is capable of accepting inputs of 100 to 240 VAC, 50/60 Hz. The monitor’s power consumption is 15 Watts. It is highly recommended that the monitor be connected directly to the AC power source, preferably on its own circuit. The connection should be completed with UL approved multi-conductor wire (1418 AWG), rated 300 VAC at 105°C.

Locate a convenient service knockout and install electrical conduit in the typical manner. Locate the AC Input Terminals on the inside of the monitor (Page 8) and secure the incoming AC power leads to the GND, LINE 1, and LINE 2 terminals as shown in the diagram below.

A switch or circuit breaker rated 1.0 A, 250 VAC, with a minimum terminal spacing of 3.0 mm must be attached to the monitor’s AC power leads. This switch must also be located in close proximity to the monitor, and be in easy reach of the operator. This switch should also be clearly marked as the monitor’s main AC disconnect device.

AC Power Connector Board

GND

LINE1

LINE2

 

NEUTRAL

 

WARNING: Drilling holes in the HGMSZ enclosure will void the warranty. Please use knockouts provided for electrical connections.

WARNING: Electrical installation should be performed by a certified electrician, and should comply with all applicable NEC/CEC and local electrical safety codes.

WARNING: Connection of the mains wiring must be made to terminals LINE1, LINE2, and GND. Under no circumstances should this monitor be operated without a protective ground. Doing so poses a potential shock hazard, and is also a violation of electrical safety standards applicable to this type of equipment.

Instruction 3015-4256

9

Image 13
Contents Hgmsz Halogen Gas Monitor Single Zone Installation / Operation / MaintenanceDeclaration of Conformity Table of Contents Entering Custom Refrigerant Calibration Values Hazard Symbols on Monitor How to Use This ManualSafety Precautions General Description Functional OverviewUnderstanding Monitoring Levels Response to the Presence of Multiple RefrigerantsSuggested Location of Sampling Point Inspection Installation ConsiderationsLocating the Monitor Mounting Instructions Connecting Gas Sample and Exhaust LinesOverview Connecting the Exhaust Line Connecting the Gas-Sample LineTubing Considerations Interior Schematic Interior SchematicElectrical Wiring Connection Connecting External AlarmsMA Current Loop Interface Typical External Alarm Relay 1 WiringInstruction General Operation Front Panel Display and ControlsDisplay Screens Initial Power UpData Display Screen Function Screens R134A Calfactor Audible Alarm is OFF Working with System Faults Working with Gas AlarmsFault Codes Viewing the Fault LogReset to Factory Default Settings Clearing the Stored PPM Log, Alarms & Faults DataEntering Custom Refrigerant Calibration Values Second Diagnostic Screen First Diagnostic ScreenWorking with the Diagnos Function Keypad FunctionsScreen Display Working with the P-CHK FunctionAdjusting Calibration Factor Calibration ProcedureWorking with the Calibration Function Instruction Charcoal Filter Clock BatteryHydrophobic Filter Servicing Air LinesReplacement Parts Replacement Parts & Optional AccessoriesOptional Accessories Appendix Table of Recommended Alarm SettingsWarranty Warranty and ServiceRegister Your Warranty by Visiting Service Field Repair Return ProcedureHalon SpecificationsPage Bacharach, Inc Dart Road, Newnan GA Ph 678-423-2481 Fax

3015-4256 specifications

Bacharach 3015-4256 is a highly regarded integrated air quality monitor designed for both commercial and residential applications. This advanced device focuses on detecting and analyzing various air pollutants, ensuring that indoor environments remain healthy and safe for occupants. Its sleek, compact design allows for easy integration into any space, making it an ideal choice for those looking to enhance air quality without compromising aesthetics.

One of the standout features of the Bacharach 3015-4256 is its advanced sensor technology. Equipped with multiple sensors, it can accurately measure levels of carbon dioxide, volatile organic compounds, particulate matter, and humidity. This multi-faceted approach to air quality monitoring ensures comprehensive data collection, allowing users to assess and understand their indoor environment better. The sensors undergo rigorous calibration to maintain accuracy over time, which is critical for effective air quality management.

Another key characteristic of the Bacharach 3015-4256 is its real-time data reporting capabilities. The device continuously monitors air quality levels and provides instant feedback through an easy-to-read display. Users can view real-time updates on air quality metrics, helping them make informed decisions about ventilation and air purification needs. Additionally, the monitor can connect to a mobile app, enabling users to access data remotely, set alerts for specific pollutant levels, and track changes over time.

Safety and user-friendliness are central to the design of this air quality monitor. The device features simple installation guidelines, requiring minimal setup to start monitoring air quality effectively. It also incorporates user-friendly controls, allowing individuals to navigate through settings and features with ease.

Moreover, the Bacharach 3015-4256 is designed with energy efficiency in mind. It operates on low power consumption, making it not only cost-effective but also environmentally friendly.

In summary, the Bacharach 3015-4256 is a sophisticated air quality monitor that offers advanced sensor technology, real-time data reporting, ease of use, and energy-efficient design. Its blend of performance and practicality makes it a valuable tool for anyone committed to maintaining a healthy indoor environment.