Specifications | 4 |
Dual Input Meter |
Dual Input Meter
The accuracy of all measurements is within ± (% of reading + number of counts) from 18 °C to 28 °C.
Add 0.1x (specific accuracy) for each °C below 18 °C or above 28 °C. For voltage measurements with 10:1 probe, add probe uncertainty +1%. More than one waveform period must be visible on the sceen.
Input A and Input B
DC Voltage (VDC) |
|
|
Ranges | 500 mV, 5V, 50V, 500V, 1250V | |
Accuracy | ±(0.5% +5 counts) | |
Normal Mode Rejection (SMR) | >60 dB | |
|
| @ 50 or 60 Hz ±1% |
Common Mode Rejection (CMRR) | >100 dB @ DC | |
| >60 dB @ 50, 60, or 400 Hz | |
Full Scale Reading | 5000 counts | |
True RMS Voltages (VAC and VAC+DC) | ||
Ranges | 500 mV, 5V, 50V, 500V, 1250V | |
Accuracy for 5 to 100% of range |
| |
DC coupled: | ............... | ±(1% +10 counts) |
DC to 60 Hz (VAC+DC) | ||
1 Hz to 60 Hz (VAC) | ±(1% +10 counts) | |
AC or DC coupled: | ±(2.5% +15 counts) | |
60 Hz to 20 kHz |
| |
20 kHz to 1 MHz | ±(5% +20 counts) | |
1 MHz to 5 MHz | ±(10% +25 counts) |
5 MHz to 12.5 MHz | ±(30% +25 counts) | |
5 MHz to 20 MHz (excl. test leads or probes) | ||
|
| ±(30% +25 counts) |
AC coupled with 1:1 (shielded) test leads | ||
60 Hz (6 Hz with 10:1 probe) | ||
50 Hz (5 Hz with 10:1 probe) | ||
33 Hz (3.3 Hz with 10:1 probe) | ||
10 Hz (1 Hz with 10:1 probe) | ||
DC Rejection (only VAC) | >50 dB | |
Common Mode Rejection (CMRR) | >100 dB @ DC | |
| >60 dB @ 50, 60, or 400 Hz | |
Full Scale Reading | 5000 counts | |
The reading is independent of any signal crest factor. | ||
Peak |
|
|
Modes | Max peak, Min peak, or | |
Ranges | 500 mV, 5V, 50V, 500V, 1250V | |
Accuracy: |
|
|
Max peak or Min peak | 5% of full scale | |
10% of full scale | ||
Full Scale Reading | 500 counts | |
Frequency (Hz) |
|
|
Ranges | 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz, |
100 kHz,1 MHz, 10 MHz, and 40 MHz Frequency Range in Continuous Autoset .....................
15 Hz (1 Hz) to 30 MHz
59