Fluke 123
Users Manual
Multimeter disturbance:
∙
∙
VDC, VAC, and VAC+DC with STL120 and short ground lead.
OHM, CONT, DIODE, and CAP with STL120, and black test lead to COM.
Table 3
Disturbance less than 1% of full scale
E= 3 V/m
E= 10 V/m
Frequency range 10 kHz to 27 MHz
VDC, VAC, VAC+DC
OHM, CONT, DIODE CAP
Frequency range 27 MHz to 1 GHz
VDC, VAC, VAC+DC
OHM, CONT, DIODE CAP
500 mV to 1250V
500Ω to 30 MΩ
50 nF to 500 μF
500 mV to 1250V
500Ω to 30 MΩ
50 nF to 500 μF
500 mV to 1250V
500Ω to 30 MΩ
50 nF to 500 μF
500 mV to 1250V
500Ω to 30 MΩ
50 nF to 500 μF
Test tool ranges not specified in table 3 may have a disturbance of more than 10% of full scale.
66