3.Insert the cartridge into the cartridge well on the front of the OnSite It should click into place when properly seated. The green PULL handle should be all the way down.

NOTE: To prevent the pads’ adhesive gel from drying out, do not open the hard cover or film seal of the cartridge until you need to use the pads.

4.Remove the battery from its packaging. Install it in the battery compartment on the back of the OnSite.

top

back

5.The OnSite will automatically run a self-test when the battery is inserted. Press the Shock button when instructed. When the self-test is over, the

OnSite will report the result, and tell you to push the green On/Off button in case of an emergency. (Do not push the green button unless this is an actual emergency.) Then the OnSite will turn off and go to standby mode. The green Ready light will be blinking to show the OnSite is ready for use.*

NOTE: Always store the OnSite with a pads cartridge and a battery installed, so it will be ready to use and can perform daily self-tests.

PHILIPS MEDICAL SYSTEMS

*As long as a battery is installed, turning the OnSite “off” puts it into standby mode, which means that it is ready for use.

2-2

Image 16
Contents A r t S t a r t O n S i t e D e f i b r i l l a t o r EditionIntentionally blank HeartStart ChirpsIntentionally blank Quick Reference Intentionally blank HeartStart OnSite Important NoteIntentionally blank Philips Electronics North America Corp Philips Medical Systems, Seattle, WA, USAIntentionally blank Contents Appendices Introduction to the Heartstart DescriptionSudden Cardiac Arrest Indications for USENational and Local Requirements Training and PracticeFor More Information Setting UP the Heartstart Onsite Package ContentsTop Back Philips Medical Systems Recommended Accessories Using the Heartstart Onsite OverviewPull the Green Handle Place the Pads Press the Shock Button Patient, and the Caution light begins flashing as a reminderTreating Infants and Children Remove the Infant/Child Pads Cartridge from its packageWhen Emergency Medical Services Arrive Do not Delay TreatmentReminders Philips Medical Systems After Using the Heartstart Onsite After Each USEOnsite Data Storage Must be replaced after being usedPhilips Medical Systems Routine Maintenance Maintaining the Heartstart OnsitePeriodic Checks RemindersTroubleshooting Tips Cleaning the OnsiteDisposing of the Onsite Accessories for the Heartstart Onsite REF 68-PCHATPhilips Medical Systems Glossary of Terms AEDCPR ECGNSA Periodic self-tests Blinking green Ready lightSee NSA pause SCAConsciousness, absence of respiration, and lack of a pulse Waveform See Smart biphasic waveformGlossary of SYMBOLS/CONTROLS Philips Medical Systems Symbol Philips Medical Systems Successful defibrillation Approved accessories are usedOnSite on someone wearing an oxygen mask Precautions Heartstart Onsite Defibrillator Specifications Technical InformationPhysical EnvironmentalCategory Sealing Controls and IndicatorsSee Electromagnetic Conformity tables Delivered Defibrillation WaveformCurrent a Shock-to-shock cycle time Disarm AED modeNewborn 14 J/kg ECG Analysis System ECG Analysis Performance Accessories Specifications Battery M5070ACategory Battery type Capacity Category Adult pads Infant/child padsAppropriate recycling facility Environmental ConsiderationsRegulations Configuration Device OptionsPatient Treatment Protocol Options NSA pause type Settings Protocol pause timer and CPRCoaching parameter settings Pause, the OnSite returns to rhythmCPR4 Installed Infants and children when an infantChild pads cartridge is installed Yes, NoPhilips Medical Systems Testing and Troubleshooting TestingTroubleshooting Recommended Action During AN EmergencyGreen Ready light is solid OnSite tells you Troubleshooting While the Onsite is in USETo insert a pads cartridge OnSite tells you To stop all motion Contact Philips for service if needed Troubleshooting While the Onsite is not in USEGreen Ready light is not on Behavior Electromagnetic Conformity Electromagnetic EmissionsEmissions test CisprElectromagnetic Immunity ESDSeparation distance according to frequency of transmitter m Environmental Considerations Important Warnings and RemindersShock Cycle Timing Intentionally blank Intentionally blank Intentionally blank Philips Medical Systems United States CanadaEurope, Middle East, and Africa Latin America
Related manuals
Manual 70 pages 12.08 Kb

M5066A specifications

The Philips M5066A is a state-of-the-art multi-parameter patient monitor designed to enhance healthcare delivery in various clinical settings. This versatile monitoring solution is widely utilized in hospitals, especially in critical care units and during transport, owing to its robust features and reliable performance.

One of the main features of the M5066A is its ability to monitor multiple vital signs simultaneously, including ECG, heart rate, invasive and non-invasive blood pressure, respiration rate, temperature, and saturation levels. This multi-parameter capability allows healthcare professionals to gain a comprehensive view of a patient’s health status, enabling timely interventions and improved patient outcomes.

The M5066A incorporates advanced technologies such as Philips’ proprietary IntelliVue technology, which enhances the accuracy and reliability of vital sign measurements. The monitor uses sophisticated algorithms to filter out noise and interferences, ensuring that the data presented to caregivers is both precise and actionable. The device also supports a range of connectivity options, allowing for seamless integration with hospital information systems and electronic medical records (EMRs). This connectivity facilitates the easy transfer and management of patient data, promoting better coordination of care.

Another significant characteristic of the M5066A is its user-friendly interface. The large, high-resolution display provides clear visuals of vital sign trends and parameters, enabling quick assessments by healthcare professionals. Customizable alarms and alerts can be programmed to suit specific patient needs, ensuring that caregivers are notified of critical changes in a timely manner.

The M5066A is also designed with patient comfort and safety in mind. The device's compact and lightweight design facilitates easy transport, while its durable construction ensures reliability in demanding environments. Furthermore, the monitor's battery backup allows for continuous monitoring even during power outages or when moving patients between departments.

In conclusion, the Philips M5066A multi-parameter patient monitor is a sophisticated tool that enhances patient monitoring by providing accurate, real-time data. Its advanced technologies, comprehensive monitoring capabilities, user-friendly interface, and focus on safety and comfort make it an invaluable asset in any clinical setting, helping healthcare providers deliver the highest standard of care.