Philips M5066A owner manual Electromagnetic Immunity, Esd

Page 62

ELECTROMAGNETIC IMMUNITY

immunity test

IEC 60601

compliance level

electromagnetic environment -

 

test level

 

guidance

 

 

 

 

electrostatic discharge

± 6 kV contact

± 6 kV contact

There are no special requirements

(ESD)

± 8 kV air

± 8 kV air

with respect to electrostatic

IEC 61000-4-2

 

 

discharge.a

power frequency

3 A/m

3 A/m

Power frequency magnetic fields

(50/60 Hz) magnetic field

 

 

should be at levels characteristic of a

IEC 61000-4-8

 

 

typical location in a typical

 

 

 

commercial/hospital environment.

 

 

 

There are no special requirements for

 

 

 

non-commercial/non-hospital

 

 

 

environments.

radiated RF

10 V/m

20 V/m

Portable and mobile RF

IEC 61000-4-3

80 MHz to 2.5

 

communications equipment should be

 

GHz

 

used no closer to any part of the

 

 

 

OnSite, including cables, than is

 

 

 

absolutely necessary.b,c The

 

 

 

recommended separation distances

 

 

 

for various transmitters and the AED

 

 

 

are shown in the following table.

 

 

 

Interference may occur in the

 

 

 

vicinity of equipment marked

 

 

 

with the following symbol:

 

 

 

 

NOTE 1. At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a.Generally, AEDs are sometimes susceptible to interference generated by patient and/or responder motion in environments in which a high static electric field is present (e.g., low humidity, synthetic carpets, etc.). As a safety measure, Philips AEDs incorporate a patented method to sense possible corruption of the ECG signal by such interference and to respond by directing the user to stop all motion. In these cases, it is important to minimize movement in the vicinity of the patient during rhythm analysis in order to ensure that the signal being analyzed accurately reflects the patient’s underlying heart rhythm.

b.The ISM (industrial, scientific and medial) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.

c.Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the OnSite is used exceeds the applicable RF compliance level above, the OnSite should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the OnSite.

PHILIPS MEDICAL SYSTEMS

H-2

Image 62
Contents A r t S t a r t O n S i t e D e f i b r i l l a t o r EditionIntentionally blank HeartStart ChirpsIntentionally blank Quick Reference Intentionally blank HeartStart OnSite Important NoteIntentionally blank Philips Electronics North America Corp Philips Medical Systems, Seattle, WA, USAIntentionally blank Contents Appendices Sudden Cardiac Arrest Introduction to the HeartstartDescription Indications for USEFor More Information Training and PracticeNational and Local Requirements Setting UP the Heartstart Onsite Package ContentsTop Back Philips Medical Systems Recommended Accessories Using the Heartstart Onsite OverviewPull the Green Handle Place the Pads Press the Shock Button Patient, and the Caution light begins flashing as a reminderTreating Infants and Children Remove the Infant/Child Pads Cartridge from its packageWhen Emergency Medical Services Arrive Do not Delay TreatmentReminders Philips Medical Systems Onsite Data Storage After Using the Heartstart OnsiteAfter Each USE Must be replaced after being usedPhilips Medical Systems Periodic Checks Routine MaintenanceMaintaining the Heartstart Onsite RemindersDisposing of the Onsite Cleaning the OnsiteTroubleshooting Tips Accessories for the Heartstart Onsite REF 68-PCHATPhilips Medical Systems CPR Glossary of TermsAED ECGNSA Periodic self-tests Blinking green Ready lightConsciousness, absence of respiration, and lack of a pulse See NSA pauseSCA Waveform See Smart biphasic waveformGlossary of SYMBOLS/CONTROLS Philips Medical Systems Symbol Philips Medical Systems OnSite on someone wearing an oxygen mask Approved accessories are usedSuccessful defibrillation Precautions Physical Heartstart Onsite Defibrillator SpecificationsTechnical Information EnvironmentalSee Electromagnetic Conformity tables Controls and IndicatorsCategory Sealing Current a Defibrillation WaveformDelivered Newborn 14 J/kg Disarm AED modeShock-to-shock cycle time ECG Analysis System ECG Analysis Performance Category Battery type Capacity Accessories SpecificationsBattery M5070A Category Adult pads Infant/child padsRegulations Environmental ConsiderationsAppropriate recycling facility Configuration Device OptionsPatient Treatment Protocol Options Coaching parameter settings NSA pause type SettingsProtocol pause timer and CPR Pause, the OnSite returns to rhythmCPR4 Child pads cartridge is installed InstalledInfants and children when an infant Yes, NoPhilips Medical Systems Troubleshooting Testing and TroubleshootingTesting Recommended Action During AN EmergencyTo insert a pads cartridge Troubleshooting While the Onsite is in USEGreen Ready light is solid OnSite tells you OnSite tells you To stop all motion Green Ready light is not on Behavior Troubleshooting While the Onsite is not in USEContact Philips for service if needed Emissions test Electromagnetic ConformityElectromagnetic Emissions CisprElectromagnetic Immunity ESDSeparation distance according to frequency of transmitter m Shock Cycle Timing Important Warnings and RemindersEnvironmental Considerations Intentionally blank Intentionally blank Intentionally blank Europe, Middle East, and Africa Philips Medical Systems United StatesCanada Latin America
Related manuals
Manual 70 pages 12.08 Kb

M5066A specifications

The Philips M5066A is a state-of-the-art multi-parameter patient monitor designed to enhance healthcare delivery in various clinical settings. This versatile monitoring solution is widely utilized in hospitals, especially in critical care units and during transport, owing to its robust features and reliable performance.

One of the main features of the M5066A is its ability to monitor multiple vital signs simultaneously, including ECG, heart rate, invasive and non-invasive blood pressure, respiration rate, temperature, and saturation levels. This multi-parameter capability allows healthcare professionals to gain a comprehensive view of a patient’s health status, enabling timely interventions and improved patient outcomes.

The M5066A incorporates advanced technologies such as Philips’ proprietary IntelliVue technology, which enhances the accuracy and reliability of vital sign measurements. The monitor uses sophisticated algorithms to filter out noise and interferences, ensuring that the data presented to caregivers is both precise and actionable. The device also supports a range of connectivity options, allowing for seamless integration with hospital information systems and electronic medical records (EMRs). This connectivity facilitates the easy transfer and management of patient data, promoting better coordination of care.

Another significant characteristic of the M5066A is its user-friendly interface. The large, high-resolution display provides clear visuals of vital sign trends and parameters, enabling quick assessments by healthcare professionals. Customizable alarms and alerts can be programmed to suit specific patient needs, ensuring that caregivers are notified of critical changes in a timely manner.

The M5066A is also designed with patient comfort and safety in mind. The device's compact and lightweight design facilitates easy transport, while its durable construction ensures reliability in demanding environments. Furthermore, the monitor's battery backup allows for continuous monitoring even during power outages or when moving patients between departments.

In conclusion, the Philips M5066A multi-parameter patient monitor is a sophisticated tool that enhances patient monitoring by providing accurate, real-time data. Its advanced technologies, comprehensive monitoring capabilities, user-friendly interface, and focus on safety and comfort make it an invaluable asset in any clinical setting, helping healthcare providers deliver the highest standard of care.