Emerson 3350, 3700 Shielded heat shrink completely covers exposed drain wires

Page 17

Shielded heat shrink completely covers exposed drain wires

g.Position gland clamping insert so the interior end is flush with the heat shrink.

h.Fold the cloth shield or braid and drain wires over the clamping insert and approximately 1/8 inch (3 mm) past the O-ring.

i.Install the gland body into the core processor housing conduit opening.

5.Insert the wires through the gland body and assemble the gland by tightening the gland nut.

6.Identify the wires in the 4-wire cable. The 4-wire cable supplied by Micro Motion consists of one pair of 18 AWG (0,75 mm2) wires (red and black), which should be used for the VDC connection, and one pair of 22 AWG (0,35 mm2) wire (green and white), which should be used for the RS-485 connection. Connect the four wires to the numbered slots on the core processor.

16

Image 17
Contents Model 3700 Transmitter MVD or Model 3350 Peripheral Before YOU Begin Installation kit European installationsChoosing a location DimensionsEnvironmental requirements Visibility of tagsDimensions face view Cable lengths Dimensions conduit openings viewMaximum cable lengths Wire remote transmitterOrienting the Model 3350/3700 optional Preparing conduit openings for Atex ZoneMounting to a flat surface Mounting the applications platformMounting the core processor Connecting input and output wiringInput/output wiring terminals Wiring terminalsConnecting the Model 3700 to the sensor Wiring terminal labelsWiring instructions for 4-wire remote installations Installation optionsWire cable to Model 3700 standard core processor Page Gland nut Shielded heat shrink completely covers exposed drain wires Reinstall and tighten the core processor lid Wire cable to core processor Connecting power supply wiring Page Page Page 20001008