Emerson Preparing conduit openings for Atex Zone, Orienting the Model 3350/3700 optional

Page 8

STEP 2. Preparing conduit openings for ATEX Zone 1

If the Model 3350/3700 carries an ATEX Zone 1 approval:

1.Remove thread protectors from conduit openings (see Figure 7).

2.Install factory-supplied cable glands or user-supplied EExe cable entry devices in conduit openings that are in use.

3.Install EExe plugs in conduit openings that are not in use.

STEP 3. Orienting the Model 3350/3700 (optional)

The Model 3350/3700 can be oriented on the mounting bracket as needed, and the display cover can be rotated on the applications platform. Figure 7 provides orientation examples.

To orient the Model 3350/3700:

1.Use the four supplied mounting bolt assemblies.

2.Using a 13 mm hex wrench, install the bolt assemblies to 12 ft-lb (16 Nm) of torque.

To rotate the display cover, if needed, see the transmitter installation manual.

Figure 7. Orientation examples

Conduit openings left

Conduit openings for

 

intrinsically safe

Display cover

sensor wiring

 

Conduit openings for

 

non-intrinsically safe

 

input/output wiring

 

Conduit openings

Conduit openings

right

down

7

Image 8
Contents Model 3700 Transmitter MVD or Model 3350 Peripheral Before YOU Begin European installations Installation kitDimensions Choosing a locationEnvironmental requirements Visibility of tagsDimensions face view Dimensions conduit openings view Cable lengthsWire remote transmitter Maximum cable lengthsPreparing conduit openings for Atex Zone Orienting the Model 3350/3700 optionalMounting the applications platform Mounting to a flat surfaceConnecting input and output wiring Mounting the core processorWiring terminals Input/output wiring terminalsWiring terminal labels Connecting the Model 3700 to the sensorInstallation options Wiring instructions for 4-wire remote installationsWire cable to Model 3700 standard core processor Page Gland nut Shielded heat shrink completely covers exposed drain wires Reinstall and tighten the core processor lid Wire cable to core processor Connecting power supply wiring Page Page Page 20001008