Sanyo MDF-594C instruction manual Lock, Control panel cover

Page 10

FREEZER COMPONENTS

1.Lock

2.Door: Hinged type. The door can be opened in any angle on the way to full open.

3.Magnetic door gasket: Seals the door and prevents leakage of cold air.

4.Inner lid: Serves as a means of reducing cold air leakage when the door is open. Remove the frost before it is accumulated too much.

5.Access port: Serves a means of leading the measuring cable from the freezing room to the outside.

6.Caster: 4 casters are provided. They make the moving of the unit easier.

7.Leveling foot: Serves to adjust the height and to settle the frame.

8.Temperature display area: The digital temperature indicator, alarm lamp, and filter check lamp are installed.

9.Exhaust air vent: Be careful not to block this.

10.Control panel cover

11.Control panel cover lock: To lock the control panel cover to avoid the setting by accidental contact.

12.Control panel: Temperature set key, power switch and so on are installed. Refer to page 10 for details.

13.Back-up system joint (rear side): It is positioned at rear of the unit. Serves to connect with the pipelines from the liquid CO2 cylinder.

14.Side panel: For cleaning the condenser filter and condenser. The remote alarm terminal is located inside the panel.

15.Air intake vent (grille): Do not block this vent to keep the proper cooling performance.

16.Space for temperature recorder: An automatic temperature recorder can be attached here. (See

page 26 to 29)

9

 

36

 

Image 10
Contents MDF-594C MDF-793C Contents Introduction Precautions for Safe Operation Unqualified personnel may cause electric shock or fire Abnormal operation may cause electric shock or fire Precautions for Safe Operation Environmental Conditions Freezer Components Lock Control panel coverControl panel Installation Site Removing the packaging materials and tapes InstallationAdjusting the leveling foot Fixing the unitOperation after power failure START-UP of UnitChamber Temperature Setting Remote Alarm TerminalAlarm Temperature Setting Alarms & Safety Functions Setting of Alarm Resume TimeAlarm and safety function RED Black White Normal defrosting Routine MaintenanceThorough defrosting Removal of nickel-cadmium battery Replacement of BatteryLocation of a nickel-cadmium battery Handling of batteryTroubleshooting Malfunction Check/RemedyRecycle of battery Disposal of UnitEnglish GermanFrench SpanishItalian PortugueseDutch SwedishFeeding of chart Replacement of cellSetting of back-up temperature PenholderControl Panel Temperature Recorder Option BACK-UP System Option Specifications PerformanceSafety check sheet JK1035-150B
Related manuals
Manual 35 pages 59.46 Kb

MDF-594C specifications

The Sanyo MDF-794C and MDF-594C are advanced laboratory refrigerators designed for the safe storage of biological samples, vaccines, and sensitive materials. These models stand out due to their robust construction, user-friendly features, and innovative technologies that cater to the requirements of research institutions, hospitals, and laboratories.

One of the main features of the Sanyo MDF-794C is its large capacity, which makes it suitable for facilities with high storage demands. The refrigerator can accommodate a significant volume of samples while maintaining a stable internal environment. The MDF-594C, while slightly smaller, still offers ample storage space without compromising on performance.

Both models incorporate advanced cooling technologies, including a reliable compressor system that ensures rapid temperature recovery even during door openings. They are capable of maintaining temperatures ranging from -10°C to -30°C, ideal for a variety of laboratory applications. Temperature uniformity is a crucial aspect of sample preservation, and Sanyo’s state-of-the-art design guarantees even distribution of cold air throughout the cabinet.

Energy efficiency is another notable characteristic of the Sanyo MDF-794C and MDF-594C. These models are designed to operate with minimal energy consumption, thereby reducing operational costs and environmental impact. Additionally, they are equipped with a reliable insulation system, which helps maintain consistent temperatures while minimizing energy loss.

User-friendliness is emphasized in the design of both refrigerators. They feature intuitive, programmable digital control panels that allow users to set and monitor temperatures easily. The alarms for temperature deviation and system malfunctions ensure that users are promptly alerted to any issues, enhancing safety and reliability.

Another significant feature is the inclusion of a self-diagnostic system in both models, which continuously monitors the performance of the unit. This capability aids in early detection and troubleshooting, thus maintaining optimal operating conditions.

The Sanyo MDF-794C and MDF-594C also prioritize hygiene and cleanliness, with easy-to-clean interiors that help prevent contamination. Their design allows for easy organization of samples, thus improving workflow within the laboratory.

In conclusion, the Sanyo MDF-794C and MDF-594C are exemplary laboratory refrigerators characterized by their capacity, advanced cooling technologies, energy efficiency, user-friendly controls, and commitment to sample safety. These attributes make them invaluable assets in any laboratory setting, ensuring the integrity and reliability of stored materials.