Emerson Process Management 396PVP Section Installation, Unpacking and Inspection, Mounting

Page 9

MODEL 396P and 396PVP

SECTION 2.0

 

INSTALLATION

SECTION 2.0

INSTALLATION

2.1 UNPACKING AND INSPECTION.

Inspect the outside of the carton for any damage. If dam- age is detected, contact the carrier immediately. Inspect the hardware. Make sure all the items in the packing list are present and in good condition. Notify the factory if any part is missing. If the sensor appears to be in satisfactory condition, proceed to Section 2.2, Mounting.

NOTE

Save the original packing cartons and materi- als as most carriers require proof of damage due to mishandling, etc. Also, if it is necessary to return the sensor to the factory, you must pack the sensor in the same manner as it was received. Refer to Section 6.0 for return instructions. If the sensor is to be stored, the vinyl boot should be filled with pH buffer solu- tion and replaced on sensor tip until ready to use.

CAUTION

Buffer solution, in the vinyl boot, may cause skin or eye irritation.

WARNING

Glass electrode must be wetted at all times (in storage and in line) to maximize sensor life.

2.2 MOUNTING.

The sensor has been designed to be located in indus- trial process environments. Temperature and pressure limitations must not be exceeded at any time. A caution label regarding this matter is attached to the sensor. Please do not remove the label. See Figure 2-1.

CAUTION

Internal electrolyte fill solution may cause skin or eye irritation.

Mounting Guidelines:

1.Shake the sensor in a downward motion to remove any air bubbles that may be present inside the tip of the pH glass.

2.Do not install the sensor on the horizontal. The sensor must be 10° off the horizontal to ensure accuracy.

3.Do not install the sensor upside down.

4.Air bubbles may become trapped in the sensor end between the glass bulb and the sensor body. This problem is most commonly encountered in areas of low flow or during calibration. Shake the probe while immersed in solution to remove bubbles. This problem can be avoided by ordering the sen- sor with the slotted tip (option -41).

In most cases, the pH sensor can simply be installed as shipped and readings with an accuracy of ± 0.6 pH may be obtained. To obtain greater accuracy or to ver- ify proper operation, the sensor must be calibrated as a loop with its compatible analyzer or transmitter.

2.2.1Flow Through and Insertion Mounting. Model 396P and 396PVP Sensors have a 1-inch MNPT process connection at the front of the sensor for mounting into a 1-1/2 inch tee or the process pipes. See Figure 2-1 through Figure 2-7 for installation con- figurations.

NOTE

LARGE PIPE WRENCHES MUST NOT BE USED TO TIGHTEN THE SENSOR INTO A FLANGE OR OTHER TYPE OF MOUNTING.

2.2.2Submersion Mounting. Model 396P and 396PVP Sensors also have a 1 inch MNPT process connection at the back of the sensor. Utilizing a stan- dard 1 inch union, the sensor may be mounted to a 1 inch SCH 80 CPVC or PVDF standpipe (see Figure 2- x). Tapered pipe threads in plastic tend to loosen after installation. It is therefore recommended that Teflon1 tape be used on the threads and that the tightness of the connection be checked frequently to assure that no loosening has occurred. To prevent rain water or con- densation from running into the sensor, a weatherproof junction box is recommended (see Figure 2-x). The sensor cable must be run through a protective conduit for isolation from electrical interference or physical abuse from the process. The sensor should be installed within 80° of vertical, with the electrode facing down. The sensor’s cable should not be run with power or control wiring.

1Teflon is a registered trademark of E.I. du Pont de Nemours & Co.

5

Image 9
Contents Combination pH/ORP Sensor Essential Instructions Hazardous Area InstallationModel 396P and 396PVP Combination pH/ORP Sensor List of Figures Features and Applications Section Description and SpecificationsPercent Linearity SpecificationsOrdering Information Recommended Accessories for First Time InstallationsPart Description Other Accessories for Models 396P and 396PVPSection Installation Unpacking and InspectionMounting 40396P01 Dimensional Drawing40396P02 40396P04 40396P03Low Flow Cell PN Wiring Matrix Guide Section Wiring Model 396P-01Wire Functions for Model 396P with and without Preamplifier 40396P15 40396P13 40396P08 Section Refer to Section Wiring Models 396P-02 and 396PVPConnector Pins Their Functions Section Section Section 20. Wiring to Model 54e pH/ORP With Variopol 40396P24 40396P2523. Wiring Model 396P-01 to Model Xmt-P-HT-10 ORP Calibration Section Start UP and CalibrationAutomatic Temperature Compensator Section MaintenanceCompensation Elements ElementsModel 396P and 396PVP ORP Diagnostic Message Section Diagnostics and TroubleshootingDescription of Problem Remedy UG!1SSTroubleshooting Without Advanced Diagnostics Description Quantity NON-WARRANTY Repair Warranty RepairSection Return of Material GeneralReturn of Material WarrantySpecifications subject to change without notice

396P, 396PVP specifications

Emerson Process Management has established itself as a leader in the automation industry, providing innovative solutions that enhance operational efficiencies and improve process control. Among its notable offerings are the 396PVP and 396P pressure transmitters, designed to meet the needs of various industrial applications.

The 396PVP model is equipped with advanced features that facilitate high-performance measurements in both challenging and standard environments. This pressure transmitter utilizes a piezoresistive sensor, allowing for accurate and reliable pressure readings. It supports a wide range of measurement capabilities, making it suitable for liquids, gases, and vapors across diverse industries. With an impressive pressure range, the 396PVP caters to various applications, from HVAC systems to more demanding scenarios in oil and gas extraction.

In terms of technology, the 396PVP leverages digital communication protocols, including HART and FOUNDATION Fieldbus, enabling seamless integration into existing control systems. This capability not only enhances data transmission but also facilitates remote monitoring and diagnostics, significantly reducing downtime and maintenance requirements. The transmitter is also designed with an intuitive interface, allowing for straightforward configuration and calibration via handheld devices or software.

On the other hand, the 396P model offers exceptional performance for demanding pressure measurement requirements. It features robust construction materials that ensure durability and resistance to harsh environments, making it an ideal choice for applications in chemical processing and power generation. The 396P is designed with a fast response time, ensuring real-time data accuracy, which is critical for maintaining process integrity and safety.

Both models showcase Emerson's commitment to safety and reliability. They are built to meet stringent industry standards, which guarantees quality performance in hazardous locations. Furthermore, the embedded diagnostic features provide real-time health monitoring, alerting operators of any potential issues before they escalate, thereby safeguarding operations.

In summary, the 396PVP and 396P pressure transmitters from Emerson Process Management exemplify cutting-edge technology and high-quality engineering. With their advanced sensing capabilities, robust design, and integration-friendly features, these transmitters are essential tools for optimizing process control and enhancing overall productivity in various industrial settings. Whether it's for precise measurements in a chemical plant or reliable pressure readings in an oil field, the 396 series offers the performance and durability required to thrive in today's demanding operational landscapes.