Bunn LCA-1, LCR-1 service manual Calibration of the Concentrate Pump & Dispenser Flow Rate

Page 21

CALIBRATION OF THE CONCENTRATE PUMP & DISPENSER FLOW RATE

The factory set default values for the Pump & Dispenser Flow Rates are very accurate and typically do not need to be field calibrated. However, if the mix ratio accuracy is ever in question, this procedure can be used to reca- librate the unit in the field.

Equipment Required:

50 ml graduated cylinder, with 1 ml graduations.

1000 ml graduated container.

NOTE: You can calibrate either the Concentrate Pump or the Dispenser Flow Rate independently. Simply scroll through the menu screen to the desired section and perform only those steps.

Calibrating the Concentrate Pumps.

1.Open the door of the dispenser and select PROGRAM on the Selector Switch. The display should show PPP.

2.Depress and Hold both Up & Down arrows switch for approximately 15 sec. The unit will display “1 run” when it has entered the Calibration Mode.

3.Select PRIME on the Selector Switch.

4.Place a container under the dispense tip and depress the dispense switch until a steady stream of concentrate comes out the tip (5 to 10 seconds).

5.Stop priming and allow the tip to stop dripping. Discard the concentrate collected.

6.Place a 50 ml graduated cylinder under the Dispense Tip.

7.Press & Hold the dispense switch. The dispenser will display a 20 second count down timer and then shut the pump OFF automatically. You can abort the calibration and stop the pump by releasing the switch.

8.Keep the graduated cylinder under dispense tip until all the concentrate has dripped out.

9.Measure the volume of concentrate collected in the graduated cylinder.

The acceptable range for the volume of concentrate collected is 16 - 26 ml

10.If the amount collected is not within the acceptable range, repeat STEPS 6 – 9.

11.If the amount collected is still not within range, replace the pump tubing with a new Tube Kit, (refer to the Tube Replacement Instructions) and restart this procedure.

12.When satisfied with the volume of concentrate collected, press the Down Arrow switch. The current Calibra- tion volume “1 XX” will be displayed.

13.Use the (-) or (+) keys to adjust the number displayed to the amount measured in STEP 9.

(Continued Next Page)

21

44806.1 081211

 

Image 21
Contents Phone 217 529-6601 FAX 217 LCR-1 LCA-1BUNN-O-MATIC Commercial Product Warranty Contents 00656.0001 User Notices27442.0000 33461.0005 11646.0002Introduction Electrical RequirementsPlumbing Requirements Locating the Dispenser Initial SET-UPPlumbing HOOK-UP Electrical HOOK-UPUser Interface Operating Controls and InterfaceOperating Controls and Interface Installing Pump Tubing Installing the Pump TubingPriming the Concentrate Line Initial Fill & Heat Rinse Alarm Feature Cabinet Target Temperature LCR-1 OnlyLevel Selected Alarm Display Mode Description Programming the DispenserLED Display Programming the Dispenser Program Mode DescriptionOperating the Dispenser Daily Rinsing General Cleaning and Sanitizing ProceduresDaily Parts Washing Weekly SanitizingCleaning the AIR Filter LCR-1 Only Preventive MaintenanceReplacing the Pump Tubing Replacing the Water FilterCalibrating the Dispenser Calibration Mode  DESCRIPTIONDefault setting Reset Factory DefaultsReset by holding the Decrease Calibrating the DispenserTroubleshooting Guide Error CodesLED Display Error Troubleshooting Guide Error Codes Follow errors codes apply to the LCR-1 onlyDESCRIPTION/TROUBLESHOOTING Calibrating the Concentrate Pumps Calibration of the Concentrate Pump & Dispenser Flow RateCalibration of the Concentrate Pump & Dispenser Flow Rate Draining the HOT Water TankCalibrating the Dispenser Flow Rates Tank Heater Schematic Wiring Diagram LCR-1 / LCA-1