VII. INTELA-TRAUL® (continued)
VII. a - ALARM EXPLANATIONS:
*NOTE: Explanation of alarms assume the audible alarm style is set at a
High Cabinet Air Temperature: The audible alarm* will sound and the display will read HI CAb when the temperature inside the cabinet rises above a
POSSIBLE CAUSES (for Refrigerator & Freezer Models):
•Doors open for extended periods of time.
•Large amounts of hot product placed inside the cabinet.
•Condenser coil dirty.
Low Cabinet Air Temperature: The audible alarm* will sound and the display will read Lo CAb when the temperature inside the cabinet falls below a
POSSIBLE CAUSES (for Refrigerator & Freezer Models):
•No product in unit.
•Failed sensors.
Loss Of Power: The audible alarm* will sound and the display will read Electrical Loss, when the unit regains power after an outage. To turn off the audible alarm* and/or clear the visual text, press the alarm cancel button. You must reset the clock and date if you are using any defrost lockouts or are retrieving the information from the data storage memory. See “Setting The
Door Open: The door open icon on the face of the control will light and the display will read Door Open, after the door or doors have been open for 2 minutes. If the door remains open for an additional 13 minutes the audible alarm* will sound. To turn off the alarm, close the door or press the alarm cancel button. If the door remains open for an additional 5 minutes, the audible alarm* will sound again (n/a on
System Leak: The audible alarm* will sound and the display will read Call Service, when the control detects a leak in the refrigeration system. To turn off the audible alarm*, press the alarm cancel button. The visual alarm text will remain until a service technician has repaired the unit. If the condition remains for 24 hours, the audible alarm* will sound again.
POSSIBLE CAUSES:
•Low refrigerant charge.
•Discharge sensor has failed low.
CONDENSERCLEAN™: The audible alarm* will sound and the display will read Clean Filter, when the operating pressures and temperatures exceed a safe operating range. As the load on the condenser decreases, the alarm will turn off by itself. As the pressures on the condenser continues to rise, the audible alarm* will stay on for longer periods of time. To turn off the audible alarm*, press the alarm cancel button. The visual alarm text will remain until the filter/condenser has been cleaned. If the condition remains for 24 hours, the audible alarm* will sound again and visual display will read Call Service.
POSSIBLE CAUSES:
•Inadequate air flow through condenser unit.
•Discharge sensor has failed high.
Sensor Failures: The audible alarm* will sound and the display will read Sensor Failure Call Service, when any of the unit sensors fail to operate. To turn off the audible alarm*, press the alarm cancel button. The visual alarm text will remain until the sensor has been replaced. Depending on the function of the sensor, the audible alarm* will sound again in either 5 minutes or 24 hours.
*Not available on remote models.