PAS Security code set up: (0 999)

tPb

Type of probe: it displays the probe used (NTC or PTC) (readable only)

rEL

Release code (readable only)

Ptb

Parameter code (readable only)

Printer Management

The instrument XB570L can drive the serial printer XB05PR by means of the RS232 output. So temperatures measured during a freezing cycle can be printed.

The internal Real Time clock of the instrument provides time reference.

The “Prt” functions manages the printer by means of the following parameters: Prt: printer presence: (Yes/no) it enables or disables the printer.

itP: printing interval: (0.0÷30.0 min).

PbP: which probe has to be printed: (iP: Only the insert probe; rP: Only the thermostat probe; irP: Insert and

thermostat probes; irE: All the probes.

PtH: Printing during the holding phase (yES, no) yES = Printing enabled during the holding phase; no = Printing disabled during the holding phase.

NOTE: Although the controller is capable of several features ie. HACCP recording and output, it is intentional by Foster that these have not been developed to maintain the simplicity of operation for the end user. Therefore these features are not offered or supported.

Alarm Signals

Mess.

Cause

Outputs

“EE”

Data or memory failure

Alarm output ON; Other outputs unchanged

“rPF”

Thermostat Probe failure

Alarm output ON; Compressor output according to parameters

COn” and “COF

 

 

“EPF”

Evaporator Probe failure

Alarm output ON; Defrost termination is timed; No temperature

control on fans.

 

 

“iPF”

Insert probe failure

Alarm output ON; Other outputs unchanged; The cycle is made by

time

 

 

“rtC”

Real Time Clock data lost

Alarm output ON; Other outputs unchanged;

“rtF”

Real Time Clock failure

Alarm output ON; Other outputs unchanged; The date and the

duration of the cycle are not available.

 

 

“HA”

Maximum temperature alarm

Alarm output ON; Other outputs unchanged.

“LA”

Minimum temperature alarm

Alarm output ON; Other outputs unchanged.

“FF”

Fast freezing interrupted by short power failure

Alarm output ON; The freezing cycle restart from the same point at

 

 

which was interrupted.

“PFA”

Fast freezing interrupted by long power failure

Alarm output ON; The freezing cycle restart from the current phase.

“OCF”

Max duration of the cycle is expired

Alarm output ON; Other outputs unchanged. In any case the cycle

ends when the final temperature is reached

 

 

“EA”

External alarm

Alarm output ON; Other outputs unchanged.

“CA”

Serious external alarm

Alarm output ON; Other outputs OFF.

“dA”

Door open alarm

Alarm output ON; Other outputs unchanged.

Controller Electrical Connections

12

Image 13 Contents
Page Contents Longer Term Storage ‘Hold’ Mode Conservation Program Hard Chilling TemperatureCook Chill Operation Program Soft ChillingOperation QC 11, QC 600 & Bqcf 40 Blast Chill Operating InstructionsManual Defrost IntroductionTemporarily Stopping the Cycle Mode FunctionIlluminated Page Bqcf 40 Cycle Parameters QC 60 Cycle ParametersFunction Menu Bqcf Parameter SettingsParameter Definitions DiC Digital input operating mode EAL, bAL, dorRPO OAt FnC Fans operating mode during the holding phaseOAS Set point for AUX output -50÷50 ris.1 C/ 1F OAiRPF Printer ManagementRefrigerant R404a Refrigerant Quantity Model Wiring Diagram Number Fuse RatingQC 11 Wiring Diagram QC 600 Wiring Diagram Bqcf 40 Wiring Diagram. Sheet 1 Bqcf 40 Wiring Diagram. Sheet2 Email commercial@fosterfrance.com