Agilent Technologies 6632B, 6634B, 66332A, 6633B service manual CV Source Effect, CV Noise Pard

Page 15

Verification and Performance Tests - 2

e.Open the load and again record the DVM voltage reading. The difference between the DVM readings in steps

(d) and (e) is the load effect voltage, and should not exceed the value listed in the performance test record chart for the appropriate model under CV LOAD EFFECT.

CV Source Effect

This test measures the change in output voltage that results from a change in ac line voltage from the minimum to maximum value within the line voltage specifications.

a.Turn off the supply and connect the ac power line through a variable voltage transformer.

b.Connect the output as shown in Figure 2-1a with the DVM connected between the +S and the -S terminals. Set the transformer to nominal line voltage.

c.Turn on the supply and program the current to the maximum programmable value and the output voltage to the full-scale value .

d.Adjust the load for the full-scale current value as indicated on the front panel display. The CV annunciator on the front panel must be on. If it is not, adjust the load so that the output current drops slightly.

e.Adjust the transformer to the lowest rated line voltage (e.g., 104 Vac for a 115 Vac nominal line voltage input).

f.Record the output voltage reading on the DVM.

g.Adjust the transformer to the highest rated line voltage (e.g., 127 Vac for 115 Vac nominal line voltage input).

h.Record the output voltage reading on the DVM. The difference between the DVM reading is steps (f) and (h) is the source effect voltage and should not exceed the value listed in the performance test record chart for the appropriate model under CV SOURCE EFFECT.

CV Noise (PARD)

Periodic and random deviations (PARD) in the output (ripple and noise) combine to produce a residual ac voltage superimposed on the dc output voltage. CV PARD is specified as the rms or peak-to-peak output voltage in the frequency range specified in the User’s Guide.

a.Turn off the supply and connect the output as shown in Figure 2-1a to an oscilloscope (ac coupled) between the

(+) and the (-) terminals. Set the oscilloscope's bandwidth limit to 20 MHz and use an RF tip on the oscilloscope probe.

b.Turn on the supply and program the current to the maximum programmable value and the output voltage to the full-scale value.

c.Adjust the load for the full-scale current value as indicated on the front panel display.

d.Note that the waveform on the oscilloscope should not exceed the peak-to-peak limits in the performance test record chart for the appropriate model under CV NOISE (PARD).

e.Disconnect the oscilloscope and connect an ac rms voltmeter in its place. The rms voltage reading should not exceed the RMS limits in the performance test record chart for the appropriate model under CV NOISE (PARD).

15

Image 15
Contents Agilent Part No Microfiche No September Certification Warranty InformationBefore Applying Power Safety SummarySymbol Description Safety Symbol DefinitionsInstrument Identification Printing HistoryTable of Contents Post-repair Calibration Initialization ROM UpgradeDisassembly Procedures Introduction General Schematic Notes BackdatingPage Chapter Safety ConsiderationsOrganization Related DocumentsFirmware Revisions RevisionsElectrostatic Discharge Manual RevisionsTest Equipment Required IntroductionMeasurement Techniques Test SetupCurrent-Monitoring Resistor Operation Verification TestsPerformance Tests Electronic LoadCV Load Effect CV SetupConstant Voltage CV Tests Voltage Programming and Readback AccuracyCV Noise Pard CV Source EffectCurrent Programming and Readback Accuracy Transient Recovery TimeCC Setup Constant Current CC TestsCurrent Sink -CC Operation Low Range Current Readback AccuracyCC Load and Line Regulation CC Source Effect CC Load EffectCC Noise Pard Performance Test Equipment FormPerformance Test Record Forms Iout Iout + 4.3 mA Iout + 0.25 mAIout Iout + 2.3 mA Troubleshooting Type Purpose Recommended Model Overall TroubleshootingFlow Charts Test Equipment Required for TroubleshootingSheet 1. Main Flowchart Sheet 2. Main Flowchart Sheet 3. Main Flowchart Sheet 4. OV at Turn-On Sheet 5. OV at Turn-On Sheet 6. FS Indicated but Fuse OK Sheet 7. No Output Voltage Sheet 8. No Output Voltage Sheet 9 No Output Voltage Sheet 10. No Current Limit Sheet 11. Unit Does Not OV Sheet 12. High Output Voltage Error Code Description Probable Cause Specific Troubleshooting ProceduresPower-on Self-test Failures Self-Test Error Codes/MessagesBias Test Point Common Measurement CV/CC Status Annunciators TroubleshootingBias and Rail Voltages Bias and Reference VoltagesVoltage Measurements at J307 A2 Interface to A1 Main board J307 Voltage MeasurementsManual Fan Speed Control Disabling Protection FeaturesInhibit Calibration Switch Calibration PasswordPost-repair Calibration Identifying the Firmware ROM UpgradeUpgrade Procedure InitializationDisassembly Procedures Remote sense leads before attempting disassemblyList of Required Tools Cover, Removal and Replacement A2 Interface Board, Removal and ReplacementFront Panel Assembly, Removal and Replacement A1 Main Control Board S1 Line Switch, Removal and ReplacementA3 Front Panel Board, Removal and Replacement T1 Power Transformer, Removal and ReplacementTransformer Wiring Line Voltage WiringInterface Signals Power Supply Interface signalsConnector Signal Description Secondary Interface A3 Front Panel CircuitsA2 Interface Circuits Primary InterfaceA2/A3 Block Diagram A1 Main Board Circuits Power CircuitsA1 Block Diagram Control Circuits Principles of Operation Page Designator Model PartNumber Qty Description Chassis, ElectricalDesignator Model Part Number Qty Description Chassis, MechanicalMechanical Parts ldentification A1 Control Board PC Board Assembly Replaceable Parts Replaceable Parts Replaceable Parts Replaceable Parts Replaceable Parts Replaceable Parts Replaceable Parts Replaceable Parts Replaceable Parts Replaceable Parts Replaceable Parts Binding Post Option #020 A5 AC input/RFI BoardRelay Option #760 Model-dependent Components General Schematic NotesDiagrams A1 Board Component Locations R434 825 R500 175 325 R435 R505 A1 Board Component Locations A4 and A6 Board Component Locations A1 Board schematic sheet A1 Board schematic sheet A1 Board schematic sheet A6 Relay Option Board schematic Index Index
Related manuals
Manual 103 pages 34 Kb Manual 83 pages 60.43 Kb

6632B, 6633B, 66332A, 6634B specifications

Agilent Technologies, a leader in electronic test and measurement solutions, offers a range of power supplies designed to meet various application needs. Notable models include the 6632B, 6634B, 66332A, 6633B, and 6612C. Each of these units provides unique features and technologies that cater to researchers, engineers, and technicians in the industry.

The Agilent 6632B is a single-output DC power supply that delivers up to 30V and 3A. It is known for its excellent load regulation and low noise, making it ideal for sensitive electronic testing. The model includes built-in voltage and current measurement capabilities, allowing users to monitor output conditions in real time. The 6632B is commonly used in laboratory environments, educational institutions, and manufacturing lines.

Moving to the 6634B, this model offers dual-output capabilities with a maximum output of 30V and 6A. This versatility enables simultaneous powering of two different devices or circuit sections. It also features parallel and series operation options, allowing users to create a custom power supply configuration for specific applications. With a programmable interface, the 6634B simplifies test automation, ensuring efficiency in extensive testing scenarios.

The Agilent 66332A stands out with its precision and high performance. This power supply provides three outputs—two programmable and one fixed—yielding flexible power configurations. Its intuitive user interface allows easy adjustment of voltage and current settings. The device is equipped with extensive protection features to safeguard both the power supply and the connected load against faults. It is an excellent choice for complex testing setups that require reliable power.

The 6633B model offers a high-performance power supply with dual outputs, similar to the 6634B but with enhanced specifications. It can provide up to 40V and 2A per channel, delivering precision for demanding applications. This model is particularly suited for industries focused on high-reliability applications, such as telecommunications and aerospace.

Lastly, the Agilent 6612C is a compact and lightweight power supply providing single-output up to 60V and 2A. This model is designed for simplicity and ease of use, making it an excellent choice for portable applications. The 6612C’s unique characteristics include a compact design and user-friendly controls, which facilitate operation in field settings.

In summary, Agilent Technologies’ power supply models—6632B, 6634B, 66332A, 6633B, and 6612C—offer an array of features that cater to a wide range of testing and research needs, ensuring reliable power delivery in various contexts.