Setting Up Generic and Advanced Signals | Setting Up a Pulse Signal |
Setting Up a Pulse Signal
Task Set up a continuous pulse signal with 20 ns period, a pulse width of 5 ns, an amplitude of 2.0 V and an offset of 1.5 V (high level 2.5 V, low level 500 mV).
Pulse Width = 5 ns
2.5 V
500 mV
Period = 20 ns
Using the Graphical User Interface
Instrument Settings To set the instrument parameters:
1Disable the outputs.
2In the instrument panel, choose the Pulse/Pattern mode.
3Set the Period to 20 ns.
Channel Settings To set the channel parameters:
1Switch to the Channel page.
2Select the pattern mode Pulse.
3In the Timing panel:
–Switch from DCyle to Width and set the Width to 5 ns.
4In the Levels panel, set:
–The level format to Custom.
–Switch from High/Low to Amplitude/Offset.
–Amplitude to 2.0 V.
–Offset to 1.5 V.