Combining Parameters for Signal Generation | Using the Agilent 81133A/81134A Pulse Generator |
Signal Modes
In Data and PRBS pattern modes, the pulse output format can be selected from RZ, R1, and NRZ. The timing of the different format is shown in the following diagram:
1 | 1 | 1 | 0 | 0 | 1 |
RZ
R1
NRZ
•RZ
Return to 0 pulse format.
On every 0 bit of the pattern, the signal remains low. On every 1 bit of the pattern, the signal goes to high and then back to low after the time specified by the pulse width or duty cycle parameter.
•R1
Return to 1 pulse format.
On every 1 bit of the pattern, the signal remains high. On every 0 bit of the pattern, the signal goes to low and then back to high after the time specified by the pulse width or duty cycle parameter.
•NRZ
A leading edge is generated for a 0
Agilent 81133A/81134A Pulse Generator User’s Guide, January 2005 | 71 |