Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual | Specifications, Approvals, & |
Certifications |
|
2.1 EPA DESIGNATION
Teledyne API’s Model T265 Chemiluminescence Ozone Analyzer received EPA approval for Designation
on any full scale range between
with any range mode (Single, Dual, or AutoRange)
at any ambient temperature in the range of 5°C to 40°C
with a TFE filter in the sample air inlet
with a sample flow rate of 500 ± 50 cm3/min (sea level)
with the dilution factor set to 1
with Temp/Press compensation ON
in accordance with the appropriate associated instrument manuals
with or without any of the following options:
Internal or external sample pump
Sample/Cal valve option
Rack mount with or without slides
Analog input option
2.2APPROVALS AND CERTIFICATIONS
The Teledyne - API Model T265 analyzer was tested and certified for Safety and Electromagnetic Compatibility (EMC). This section presents the compliance statements for those requirements and directives.
2.2.1 Safety
IEC
CE: 2006/95/EC,
North American:
cNEMKO (Canada):
2.2.2 EMC
EN
EN 55011 (CISPR 11), Group 1, Class A Emissions
FCC 47 CFR Part 15B, Class A Emissions
CE: 2004/108/EC, Electromagnetic Compatibility Directive
2.3 WARRANTY
Refer to the Model T200 operation manual regarding the Warranty policy for the Model T265.
07337B DCN6324
11