Getting Started | Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual |
|
|
3.3 CONNECTIONS AND SETUP
This section presents connection information for setting up the instrument to operate.
3.3.1Electrical and Pneumatic Connections
1.Refer to Figure
2.Mount the NOx scrubber with base down (arrow labeled “FLOW” pointing down) onto the rear panel.
3.Connect tubing from NOx scrubber base fitting to rear panel fitting labeled “FROM SCRUBBER”.
4.Connect tubing from NOx scrubber top fitting to rear panel fitting labeled “TO SCRUBBER”.
5.Vent the EXHAUST port to atmospheric pressure and out of the room, because of its nitric oxide content.
CAUTION
GENERAL SAFETY HAZARD
Analyzer Exhaust Contains Nitric Oxide Gas.
Vent pump exhaust to a
6.Attach the sample inlet line to the sample inlet port. The pressure of the sample gas at the inlet port should be at ambient pressure and constant.
7.Using 1/8” stainless steel tubing, attach a cylinder with 10,000 ppm ±10% of nitric oxide (NO) in nitrogen (N2), with an appropriate pressure regulator, to the 1/8” stainless steel port (labeled NITRIC OXIDE) on the rear panel as shown in Figure
8.Attach a strip chart recorder and/or
9.Connect the power cord to the correct line voltage.
WARNING
Lethal voltages present inside the analyzer’s case.
Do not operate with cover off during normal operation.
Before operation, check for correct input voltage and frequency.
Do not operate without proper chassis grounding.
Do not defeat the ground wire on power plug.
Turn off analyzer power before disconnecting electrical subassemblies.