Teledyne T265 manual Step, Action, Comment

Page 24
Manual background

Getting StartedModel T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual

 

Step

Action

Comment

 

6.

Press EXIT,

The Model T265 returns to the CAL menu. Now allow

 

 

input Span gas

span gas to flow passed the instrument.

 

 

 

 

 

7.

Wait 30 min

Wait for the O3 reading to stabilize at the span value

 

 

 

(watch O3 STB).

 

8.

Press SPAN

The SPAN button should be displayed once the

 

 

 

concentration approaches the span value. In certain

 

 

 

circumstances at low span concentrations both the

 

 

 

ZERO and SPAN buttons will appear. Do not press

 

 

 

ZERO again when running span gas!

 

 

 

 

 

9.

Press ENTR

Pressing ENTR to span the instrument actually changes

 

 

 

the equations so that the concentration displayed is the

 

 

 

same as the expected span concentration you entered

 

 

 

above, thus spanning the instrument.

 

 

 

 

 

10.

Press EXIT

Pressing EXIT returns the instrument to SAMPLE

 

 

 

mode.

 

 

 

 

IMPORTANT

IMPACT ON READINGS OR DATA

Over time PMTs lose sensitivity, which is reflected by an increase in the O3 slope test function during ongoing calibrations. If the slope falls outside the range of 0.7 to 1.3, a hardware adjustment to the PMT gain is required. The process involves the adjustment of both a high and low gain potentiometer, located on the preamp printed circuit board, while noting the “NORM PMT” test function value, for a particular concentration of Ozone span gas. Typically, the high voltage to the PMT is adjusted, such that, the “NORM PMT” is twice the measured O3 concentration. See PMT Sensor

Hardware Calibration in the Model T200 manual’s Troubleshooting &Service section for detailed procedures.

Step 3 - Review Quality of calibration: Calibration Quality Check Procedure

Step

Action

 

 

1.Scroll the TEST function menu until the O3 SLOPE is displayed.

2.Scroll the TEST function menu until the O3 OFFS is displayed.

Comment

The SLOPE value for O3 should be 1.0 ± 0.3. If the SLOPE value is in the acceptable range the instrument will perform optimally.

The Model T265 will display the offset parameter “O3 OFF” for the O3 equation. This number should be near zero. A value of ±10 indicates calibration in the optimal range.

The Model T265 is now ready to measure sample gas.

22

07337B DCN6324

Image 24
Contents Manual Addendum TELEDYNE ADVANCED POLLUTION INSTRUMENTATION9480 CARROLL PARK DRIVE SAN DIEGO, CA USA Model T265 Chemiluminescence Ozone AnalyzerPage IMPORTANT SAFETY INFORMATION Customer Service DepartmentEmail: api-customerservice@teledyne.com PhoneCONSIGNES DE SÉCURITÉ ABOUT THIS MANUAL REVISION HISTORY2011 December 07, Model T265 Addendum Description of ChangesThis page intentionally left blank TABLE OF CONTENTS This page intentionally left blank 1. INTRODUCTION 1.1 OVERVIEWIntroduction 2.SPECIFICATIONS, APPROVALS, & CERTIFICATIONS Table 2-1.Model T265 Specifications 2.3 WARRANTY 2.2.1 Safety2.1 EPA DESIGNATION 2.2APPROVALS AND CERTIFICATIONSSpecifications, Approvals, & Certifications COULD DAMAGE INSTRUMENT AND VOID WARRANTY 3. GETTING STARTED3.1 UNPACKING Lift the cover straight up 3.2 INSTRUMENT LAYOUT 3.2.1 Front PanelFigure 3-1.Front Panel 3.2.2 Rear Panel Figure 3-2.Rear Panel3.2.3 Internal Chassis Layout Figure 3-3.Internal Chassis Layout3.3 CONNECTIONS AND SETUP 3.3.1Electrical and Pneumatic ConnectionsFigure 3-4.T265 Pneumatics, Basic Configuration 3.4.1 Startup 3.4.2Initial Calibrationcomponents to heat to operating temperature Step ActionComment Step 4. MAINTENANCE 4.1 MAINTENANCE SCHEDULE4.2MAINTENANCE ITEMS SPECIFIC TO THE MODEL T265 Maintenance 5. PRINCIPLES OF OPERATION NO  O3  NO2*  O2NO2*  NO2  hv Principles of Operation APPENDIX A Models T265 and 265E Menu Trees This page intentionally left blank 1Only appears when warning messages are active · · · · · · · · · · · · · · · · ·APPENDIX A-1 Software Menu Trees Sample Menu Press to cycle through the active warning07337B DCN6324 APPENDIX B - Spare Parts This page intentionally left blank T265 Spare Parts List PARTNUMBERDESCRIPTION T265 Spare Parts List Reference 07388 06/01/2011PARTNUMBER DESCRIPTIONT265 Spare Parts List Reference 07388 06/01/2011, 15PARTNUMBER DESCRIPTIONThis page intentionally left blank Appendix C Warranty/Repair Questionnaire M265E and T265Values are in the Signal I/O EMAIL api-customerservice@teledyne.comWarranty/Repair Questionnaire M265E and T265 Appendix CAPPENDIX D - Wire List and Electronic Schematics This page intentionally left blank T265 INTERCONNECT LIST T265 INTERCONNECT LIST 07337B DCN6324T265 INTERCONNECT LIST 07337B DCN6324T265 INTERCONNECT LIST 07337B DCN632407337B DCN6324