4220 Flow Meter

Section 5 Options and Accessories

Table 5-2 Multiple Analog Output Board Specifications

Operating Temperature

0 to 140° F (–18 to 60° C)

 

 

Output Accuracy

±0.5% of full-scale

 

 

Resolution

0.1% of full-scale (0-20 mA)

 

 

Electrical Isolation

Isolated from each other and from the flow meter.

 

 

Calibration

Factory-calibrated; no further adjustments necessary.

 

 

Maximum output distance

1500 feet (457.3 m) using 18 AWG wire.

 

 

Current Range (per loop)

0 to 20 mA

 

 

Maximum Load (per loop)

750 ohms

 

 

Analog Output 1 Terminations

Pin A (Red wire –)

 

Pin C (White wire +)

 

 

Analog Output 2 Terminations

Pin D (Brown wire –)

 

Pin F (Blue wire +)

 

 

Analog Output 3 Terminations

Pin E (Black wire –)

 

Pin B (Green wire +)

 

 

Note: If you must strip the cable further back to facilitate wiring, you will expose an orange, yellow, and purple wire. You may disregard these wires, as they are not connected in this application.

5.4Tipping Bucket Rain Gauge

A Tipping Bucket Rain Gauge is available from Teledyne Isco for use with 4200 Series Flow Meters. The gauge connects to the flow meter by a cable terminated in an M/S connector. This con- nector plugs into the Remote Printer/Rain Gauge connector on the case. A Y-connect cable is available that allows use of both the rain gauge and a YSI 600 Multi-Parameter Sonde at the same time. You can connect extra cable (user-supplied) between the rain gauge and the factory-supplied cable as long as you do not exceed a maximum total length of 1,000 feet. Use 18 AWG wire or larger. Do not cut the M/S connector from the cable. The rain gauge is factory-calibrated to provide a contact closure to the flow meter for every hundredth of an inch (0.01") [or 0.1 mm] of rainfall. Software in the flow meter stores this infor- mation in memory and prints the accumulated data as a text line on the printer chart. It also allows the printout to be expressed in metric. You may, if you wish, provide a rain gauge of your own choice, but to ensure compatibility with the flow meter's oper- ating environment, it must conform to the following specification:

It must provide an isolated, dry contact closure.

It must be a normally open contact configuration.

It must close for 100 ms (–25 ms +150 ms) with less than 2 ms contact bounce.

5-7

Page 115
Image 115
Teledyne 4220 installation and operation guide Multiple Analog Output Board Specifications, Tipping Bucket Rain Gauge