4220 Flow Meter

Section 1 Introduction

Table 1-2 Technical Specifications for the 4220 Flow Meter (Continued)

 

Flowlink Data Storage and Retrieval System

Memory Partitions

 

Maximum of 6 user-defined memory partitions for level or event stor-

 

 

age.

 

 

 

Data Storage

 

Rate of data storage user-selected in 1, 2, 5, 10, 15, 30, 60, or 120

 

 

minute intervals.

 

 

 

Baud Rates

 

Serial connection - 300, 1200, 2400, 4800, or 9600 baud.

 

 

Serial connection with the optional internal modem - 2400 baud.

 

 

 

Storage

 

80,000 bytes, apportioned per reading as follows: flow = 4 bytes,

 

 

level = 2 bytes, sample = 4 bytes, pH or D. O. = 1 byte)

 

 

 

Level Data

 

Level readings are stored as a 16-bit number representing .1mm

 

 

(0.0394 inch); effective range is 0–65279 meters.

 

 

 

 

 

Flow Conversion

 

 

 

Flow Rate Calculations

The flow meter creates a table of level-versus-flow rate from program selections; this table divides the level span into 256 equally-spaced level increments. Each level increment corresponds to a specific flow rate. During data collection, if the measured level falls between two table values, the flow meter will perform a linear interpolation to calcu- late the flow rate value.

Table 1-3 Technical Specifications for the 3222 Submerged Probe

 

Physical Specifications

Dimensions

78” diameter 912" long (2.2 24.1 cm).

Weight

3 lb, (including amplifier box and cable) (1.4 kg).

 

 

Body Material

CPVC (chlorinated polyvinyl chloride) housing, stainless steel.

 

 

Cable Length

25 ft. (7.6 m).

 

 

Operating Temperature

32° to 160°F (0° to 71°C).

 

 

Storage Temperature

–40° to 160°F (–40° to 71°C).

 

 

Maximum Distances

75 feet (22.7 m) with optional extension cables.

(between flow meter and level sen-

1,000 feet (305 m) with optional Quick-Disconnect Box.

sor)

 

 

 

 

Level Measurement Specifications

Measurement Range

0.1 to 10.0 ft. (0.03 to 3.05 meters).

 

 

Maximum Level

20.0 ft. (6.1 m).

 

 

Measurement Accuracy

Standard Sensor:

 

0.033 to 5.0 ft: ±0.008 ft/ft (0.01 to 1.52 m: ±0.008 m/m)

 

>5.0 ft: ±0.012 ft/ft (>1.52 m: ±0.012 m/m)

 

@ 77° F (25° C). Includes non-linearity, repeatability, and hysteresis, but does not

 

include temperature coefficient.

 

Specifications include nonlinearity, repeatability, and hysteresis, but do not

 

include a temperature coefficient.

 

 

1-9

Page 23
Image 23
Teledyne 4220 Technical Specifications for the 3222 Submerged Probe, Flowlink Data Storage and Retrieval System