Section 6 Maintenance and Service

6.2Reactivation of the Desiccators

The 4220 has a reusable desiccant canister held by a steel clamp on the inside of the case lid. There is also a tubular desiccant car- tridge on the top of the case next to the connectors. The canister contains silica gel that adsorbs moisture trapped inside the flow meter's case when it is closed. This keeps the inside of the case completely dry during shipment, storage and use. If you leave the case open, the desiccant will quickly absorb moisture from the surrounding air and will soon be saturated. It will no longer be able to protect the flow meter. Both desiccators use a color indicator that changes from blue to pink or from yellow to green when saturated. The external desiccant cartridge vents the ref- erence port of the pressure transducer.

6.2.1 Regenerating the

Look at the desiccant canister each time you open the case. The

Internal Case

canister has a window on its side that appears blue or yellow

Desiccant

when the desiccant is dry. As the desiccant absorbs moisture, the

 

window will turn pale pink or green. When the window is pink or

 

green, you need to regenerate the desiccant, or replace it with the

 

spare canister provided in the flow meter accessory package.

 

Remove the canister from the flow meter by pulling outward on

 

the spring clamp, releasing its hold on the canister.

 

Remove the canister and heat it in a vented oven in a well-venti-

 

lated room at 300°F (150°C) for about three hours, or until the

 

blue color returns. Do not use a microwave oven; the metal case

 

of the canister could cause arcing. After cooling, reinstall the can-

 

ister in the flow meter. Make sure the window on the side of the

 

canister remains visible.

Figure 6-1 Location of the Desiccant Canister

6-2

Page 136
Image 136
Teledyne 4220 installation and operation guide Reactivation of the Desiccators, Regenerating, Internal Case, Desiccant