4220 Flow Meter
Section 3 Installation
3.3.2Location of the Flow Because the flow meter uses a submerged probe for level mea-
Meter | surement, you do not have to install it directly above the primary |
| measuring device, or even particularly close to the flow stream. |
| You can install the flow meter in a convenient, protected location |
| and route the submerged probe cable to the place where the level |
| sensor is mounted. |
| For example, you can install the flow meter above ground for pro- |
| tection and easy accessibility, and then you only need to enter the |
| manhole once, for submerged probe installation and calibration |
| You will need to mount the unit within 25 feet (7.6 m), if you |
| connect the probe directly to the flow meter. |
| You can extend the distance to 50 feet (15.3 m) if you use the |
| optional 25 Foot Extension Cable. Using the optional 50 Foot |
| Extension Cable lets you increase the distance to 75 feet (22.8 |
| m). |
| For distances up to 1,000 feet, (304.8 m) use the optional |
| |
| recommended. Details on the extension cables and |
| connect Box can be found in Section 5 of this section. |
3.3.3Mounting the 4220 The 4220 does not have any special requirements for mounting. You can locate it on any relatively flat surface either horizontally, supported by the two mounting pads and the stainless steel mounting bracket, or vertically, supported by the two plastic rails on the bottom of the case.
You can also
3.3.4Suspension Above the If you suspend the flow meter above the flow stream and there is
Flow Stream | any possibility of accidental submersion, attach several feet of |
| vinyl tubing to the open vent on the end of the external desiccant |
| cartridge. You should route this tubing to a place higher than the |
| maximum possible level of the flow stream. This will prevent any |
| water from entering the vent tube for the submerged probe in |
| case of accidental submersion of the flow meter. Vinyl tubing can |
| be purchased locally or from Teledyne Isco. |
3.4
You can install the flow meter as far as 1,000 feet from the sub- merged probe if you use the