4220 Flow Meter
Section 2 Programming
Absolute barometric pressure is barometric pressure not corrected to sea level. The barometric pressure published by the U.S. Weather Bureau is adjusted to sea level. If you use their value, you must convert it to the absolute pressure for your altitude. You should use Weather Bureau barometric pressure only if you are at sea level, or are able to correct the Weather Bureau figure to absolute pressure at your location. Enter the correct value.
WRAP D. O. PROBE IN MOIST CLOTH
PRESS ENTER WHEN STABLE: X.XX MS/CM:
Then the display will advance to the following:
CALIBRATING...
PLEASE WAIT...
If you select ALTITUDE for D.O., this will appear:
UNITS FOR ALTITUDE ENTRY
• FT • • M •
Select the appropriate units and press Enter.
ALTITUDE
ALTITUDE = X.XX FT (or meters)
Enter the altitude for your location. Then:
WRAP D. O. PROBE IN MOIST CLOTH
PRESS ENTER WHEN STABLE: X.XXX MG/L
For more detailed information on the pH and D. O. probes, see
Section 5.
Returning to step 3, Parameter to Adjust, the following display will appear:
PARAMETER TO ADJUST
• NONE • • (LEVEL) • • (pH) • • (DO) • • (YSI 600) •
Note
If you are using the YSI Sonde and YSI 600 does not appear on your display, return to step 1, Program, and make sure you have selected YES from the YSI CONNECTED menu.
If you select YSI 600, the following will appear:
YSI 600 PARAMETER TO CALIBRATE
• NONE • • pH • • DO • • CONDUCTIVITY •