
4220 Flow Meter
Section 2 Programming
These devices require periodic calibration for accurate sensing. See Section 5 for more information about the dissolved oxygen probe.
If you are not using the D. O. probe, you would select NOT MEA- SURED for this step.
Selection of PPM or MG/L will keep references to pH from showing up on subsequent menus.
TEMPERATURE UNITS
• DEG F • • DEG C • • (NOT MEASURED) •
This step sets up measurement of the temperature of the flow stream. You must have an Isco Temperature Probe (or approved equivalent) attached to the flow meter's Parameter Port. The temperature probe contains a thermistor and needs no further calibration. Measurement is in degrees Celsius or degrees Fahr- enheit. If you are not using the temperature probe, you would select NOT MEASURED for this step.
Note that if you are using either the pH or D. O. probe, temper- ature must be measured; the NOT MEASURED option will not even appear.
YSI Sonde – The following series of menus concerns the use of the YSI 600
Note
4220 Flow Meters with
You can use both the YSI Sonde and a Rain Gauge on flow meters that support the YSI with a special
YSI 600 CONNECTED
• YES • • NO •
•If you select NO, you will be unable to activate the YSI Sonde later in the program.
If no communication has been confirmed, the following display will appear:
YSI COMMUNICATIONS CHECK
• YES • • NO •