4220 Flow Meter
Section 5 Options and Accessories
Greasy substances, being nonconductors of electricity, weaken the electrical potentials formed between the glass
The pH Sensor operates satisfactorily mounted either horizon- tally or vertically in the stream. However, horizontal mounting is more secure, and presents less of a debris trap.
•Simple suspension of the probe is not recommended, particularly in streams of high velocity, or those that carry debris.
If you mount the pH probe vertically, mount it securely. Do not just hang it from the top of the pipe by its cable. Suspending the probe in the stream is not stable.
For horizontal mounting, the probe fastens to a sensor carrier that
The mounting rings are held in place by the outward force of spring pressure in the smaller sizes and by a screw arrangement in the larger sizes. After mounting the probe in the ring or strap, route the cable out of the stream so it will not trap debris that could clog the sewer.
Mount the pH Module within 25 feet of the probe, but in a location higher than the highest anticipated level for the flow stream. The amplifier box is sealed and will withstand tem- porary submersion, but this should be avoided.
Note
Do not connect an Isco pH probe and temperature probe to the 201 module at the same time. The “Temperature Probe” input on the 201 module is for use ONLY with