4220 Flow Meter
Section 2 Programming
The port is not dedicated to a particular sensor, except through programming. You can change the sensor. For example, you can change from a pH probe to a D.O. sensor if you change the pro- gramming. Selection of one parameter will keep the other from showing up on the menus. Note, however, that it is possible to measure several different stream conditions including pH and D.O. at the same time with the YSI 600 Sonde.
The YSI 600 Sonde
The YSI 600 Sonde is a
The YSI 600 Sonde can measure the following water qualities: dissolved oxygen (D.O.), conductivity, temperature, and pH. Conductivity measurements made by the sonde can be used to calculate specific conductivity, salinity, and total dissolved solids. A brief description and specifications for the YSI 600 are printed in Section 5. You may also contact the factory or your Teledyne Isco representative. More information on the sonde is found in the YSI 600 Manual, shipped with each YSI 600 Sonde.
Assembled Sonde
Conductivity (inside) pH glass sensor
pH reference
|
|
| D.O. sensor | |
(Cover Removed) | Temperature | |||
|
Figure 2-1 The YSI 600 Sonde
2.3.4Step 4 - Reset Totalizer In this step, you decide whether to reset the flow meter's internal flow totalizers. If the installation is permanent, you generally won't reset the totalizer. If you are using the flow meter as a por- table recording unit and move it from site to site, you would gen- erally reset the totalizer between sites.
2.3.5 Step 5 - Sampler | It is common to use a flow meter with an Isco Automatic Waste- |
Pacing | water Sampler. Typically the flow meter signals the sampler to |
| take a sample after a certain volume has passed. It might also |
| occur after a certain condition or set of conditions has either |
| changed or been met. This step allows you to determine that |