4220 Flow Meter
Section 2 Programming
or two current loops supplied. If you use both the external con- verter and the internal board, you can have a total of four current loops controlled by the same flow meter. The internal current loops are brought out to a
To activate the internal analog output(s), return to the main screen and press 4, 2, 0.
ANALOG CAPABLE (‘0’ TO DISABLE)
OUTPUTS 3❚ (‘EXIT PROGRAM’ WHEN DONE
Additional information for the internal analog output board, including specifications for the loops, is found in Section 5.3.2.
CAUTION
Each
Use with battery powered flow meters only if:
•The battery is continuously on charge (for example with a Solar Panel Battery Charger).
•The battery is very large, such as a
•Use only one
•Keep in mind that programming choices also affect power consumption. Use “minimum” settings on the flow meter wher- ever possible. (See Section 1, Table
Even with these circumstances, you may expect significantly shorter charge life from your battery. To determine the effect of this extra current draw on battery life, please refer to the sec- tion How to Make Battery Calculations, at the end of Section 1.
The following menus determine the behavior of the
ANALOG OUTPUT
• EXTERNAL
CAUTION
If you do not have the proper hardware installed and you press 4 - 2 - 0, and the number of analog outputs is not zero, the external