
4220 Flow Meter
Section 2 Programming
The following will appear if you are measuring D. O:
DO ENABLE/ALARM
HYSTERESIS X.XXX PPM
| (or mg/l depending on units selected in Program.) | |
| A set of menus similar to those shown above for pH, D.O., and | |
| temperature will then appear for the YSI 600 Sonde, if you have | |
| selected it. You can set hysteresis for | |
| ductivity, and | |
| turned on in previous program selections. | |
2.4.2 Optional Outputs | After all the HYSTERESIS menus have been set, press Enter. | |
| The display will return to the Setup menu. This time select | |
| OPTIONAL OUTPUTS with the arrow key. Press Enter. | |
|
|
|
| OPTIONAL OUTPUTS |
|
| • ANALOG OUTPUT • • SERIAL OUTPUT • • ALARM BOX • |
|
| ALARM BOX – refers to an external accessory used to signal | |
| alarms from flow meter measurements. See Section 5 for more | |
| information about the Alarm Box. Note that choice of SERIAL | |
| OUTPUT will eliminate ALARM BOX as an option. Likewise, | |
| choice of ALARM BOX will eliminate SERIAL OUTPUT as an | |
| option. | |
| If you select any of these OPTIONAL OUTPUTS, the flow meter | |
| will request that you turn them on or off. If you are running on | |
| battery, select OFF for all unused outputs. | |
| ANALOG OUTPUT – refers to the flow meter’s capability of | |
| managing associated equipment through a | |
| The | |
| industrial processes that are variable (rather than just fully off | |
| or on). At the lower value (4 mA) the control is turned off (0%); at | |
| 20 mA the control is completely turned on (100%). In between, | |
| rates range from 1 to 99%. A typical application is a chlorinator, | |
| which must vary in application of the chlorine gas as the amount | |
| of water passing through the system increases or decreases. | |
| Current ranges other than | |
| they are less common than | |
| ported by the flow meter on the internal card only) and for longer | |
| current loops, | |
| Teledyne Isco offers two different arrangements for the | |
| control circuit. You can have either or both with the same flow | |
| meter. One arrangement requires the use of an external | |
| accessory, the | |
| module connects to the flow meter and a source of AC power and | |
| contains the circuitry necessary to create the | |
| loop. This accessory connects to the flow meter through the | |
| Interrogator connector. | |
| The other | |
| meter that contains circuitry for up to three separate, isolated | |
|