
4220 Flow Meter
Section 5 Options and Accessories
pH measurement is affected significantly by temperature, like any other chemical reaction. Consequently, temperature compen- sation is provided. in the flow meter. The Isco pH probe has a
The pH probe connects to a Parameter Module that plugs into the Parameter Port on the flow meter. The probe has a 25 ft. (7.6
m)cable. For greater distances, contact the factory. The maximum distance between the module and the flow meter is 1,000 ft. (304.8 m).
When the
The potential, when read by a sensitive voltmeter, translates into a reading of pH. With an Isco flow meter, the voltage is sent first to a preamplifier inside the probe to reduce the impedance of the circuit and improve the
5.9.1pH Probe Calibration The 4220 provides a two- or
Note
For pH probe calibration, Teledyne Isco recommends that you use a glass container for the buffer solutions to ensure that the following conditions are met:
The probe must be properly submerged in solution, and there must be no air trapped under the probe membrane, or the reading may become incorrect and/or erratic.
Do not touch the probe until after you have performed step 6.
1.To calibrate the pH probe with the flow meter, first go to program Step 1: PROGRAM, SETUP. Select PROGRAM.
2.Step through the units of measure with the Enter key until you reach the menu that says pH UNITS - NOT MEASURED, pH. Select pH. You must do this in step 1 or the pH menu will not appear later.