4220 Flow Meter

Section 5 Options and Accessories

 

 

probe. Use only 2-mil membranes with D.O. probes

 

 

connected to Isco flow meters.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 5-4 D.O. Probe Specifications

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Cathode

 

 

 

 

 

 

 

Gold

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Anode

 

 

 

 

 

 

 

Silver

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Membrane

 

 

 

 

 

 

 

FEP Teflon; 2 mil standard

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Electrolyte

 

 

 

 

 

 

 

Half-saturated KCl (Potassium Chloride)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temp. Range

 

 

 

 

 

 

 

0° to +45° C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Polarizing Voltage

 

 

 

 

 

 

 

0.8 Volts (Nominal)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Probe Current in Air at 30° C

 

 

 

 

 

 

 

19 microamps (nominal)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Probe Current in Nitrogen at 30° C

 

 

 

 

 

 

 

0.15 microamps or less

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Connection

 

 

 

 

 

 

 

12 ft. (3.2 m) cable with 5-pin male M/S connec-

 

 

 

 

 

 

 

 

 

 

tor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Response Time

 

 

 

 

 

 

 

Typical response for dissolved oxygen, using

 

 

 

 

 

 

 

 

 

 

supplied membranes, is 90% in 20 seconds.

 

 

 

 

 

 

 

 

 

 

Response at low dissolved oxygen levels is typ-

 

 

 

 

 

 

 

 

 

 

ically 90% in 60 seconds.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Figure 5-6 D.O. Parameter Module

5.10.6Calibrating the D.O. Probe with a Flow Meter

You must use the Isco 270 D.O. Module box between the probe and flow meter; this extends the distance between the probe and the flow meter to 1,000 feet.

Prepare the probe as described above if this has not already been done (fill the probe cavity with electrolyte and seat the mem- brane).

5-22

Page 130
Image 130