TROUBLE SHOOTING

If the meter locks on a particular display, unplug and plug in again the meter to unlock the display.

A total reset defaults all scaling values to zero. To perform this reset, re-apply power with RESET key pressed in.

Factory preset values for Current, Voltage and millivolt, respectively, can be programmed by simply re- apply power with either ^v, RATE or TARE key pressed in. Respective default values are for CURRENT, VOLTAGE and MILLIVOLT inputs are:

Factory default settings for voltage, current and millivolt:

 

CURRENT

VOLTAGE

MILLIVOLT

PARAMETER

INPUT

INPUT

INPUT

HIGH SCALE

20.000

10.000

100.00

LOW SCALE

0.000

0.000

0.000

OFFSET

0.00

0.00

0.00

TARE

0.000

0.000

0.000

LIMITS

1.000

1.000

1.000

Error Messages:

Various error messages indicated on the display are interpreted as below:

ERROR MESSAGES AND SOLUTIONS

ERROR MESSAGE

ERROR CONDITION

'EE EROR'

The unit was unable to

 

save programmed value

 

in EEPROM -- possible

 

failure of EEPROM.

'PIO ERR'

Analog to digital conv-

 

erter error or keyboard

 

processing error.

SOLUTION

If the message keeps coming up, unplug the unit and reapply power. If the unit still displays the error message, send it in for repair.

If the error message followed by

8.8.8.8.8.8.and the revision number are displayed occasionally, it is an indication that the power line is very noisy and needs a power line filter.

See above