TROUBLE SHOOTING
If the meter locks on a particular display, unplug and plug in again the meter to unlock the display.
A total reset defaults all scaling values to zero. To perform this reset,
Factory preset values for Current, Voltage and millivolt, respectively, can be programmed by simply re- apply power with either ^v, RATE or TARE key pressed in. Respective default values are for CURRENT, VOLTAGE and MILLIVOLT inputs are:
Factory default settings for voltage, current and millivolt:
| CURRENT | VOLTAGE | MILLIVOLT |
PARAMETER | INPUT | INPUT | INPUT |
HIGH SCALE | 20.000 | 10.000 | 100.00 |
LOW SCALE | 0.000 | 0.000 | 0.000 |
OFFSET | 0.00 | 0.00 | 0.00 |
TARE | 0.000 | 0.000 | 0.000 |
LIMITS | 1.000 | 1.000 | 1.000 |
Error Messages:
Various error messages indicated on the display are interpreted as below:
ERROR MESSAGES AND SOLUTIONS
ERROR MESSAGE | ERROR CONDITION |
'EE EROR' | The unit was unable to |
| save programmed value |
| in EEPROM |
| failure of EEPROM. |
'PIO ERR' | Analog to digital conv- |
| erter error or keyboard |
| processing error. |
SOLUTION
If the message keeps coming up, unplug the unit and reapply power. If the unit still displays the error message, send it in for repair.
If the error message followed by
8.8.8.8.8.8.and the revision number are displayed occasionally, it is an indication that the power line is very noisy and needs a power line filter.
See above
17