CARETroubleshooting& CLEANING
Please refer to page 20 for troubleshooting the installation of the Treatment Head.
The Issue | The Problem | The Solution | |
The audio signals: 5 beeps | Treatment head may not be | Properly and fully attach | |
The Information Display signals: | |||
attached, or the Treatment | |||
| treatment head. | ||
| head is defective | ||
|
| ||
The audio signals: 3 beeps |
|
| |
The Treatment Head Indicator Light displays: |
|
| |
| The treatment head |
| |
| has no remaining | Replace the treatment head. | |
| treatment cycles |
| |
The Information Display signals: |
|
|
The treatment
cycle will not start
The audio signals: 3 beeps
The Treatment Head Indicator Light displays:
|
| The battery is low | Recharge the battery. | |
| The Information Display signals: |
|
| |
| The audio signals: 5 fast beeps |
|
| |
| The Treatment Head Indicator Light displays: | The device is too | Set the device aside until it | |
|
| cold to operate | reaches room temperature. | |
| The audio signals: 10 beeps | The device is too | Set the device aside until it | |
| The Information Display signals: | |||
|
| hot to operate | reaches room temperature. | |
Battery will not hold a charge | The battery needs to | Order a replacement battery. | ||
be replaced | ||||
|
|
| ||
| Device will not turn on | The battery is dead | Charge battery | |
| The audio signals: 10 beeps |
|
| |
Device shuts off | The Information Display signals: | The device needs | Set the device aside | |
in cycle |
| to cool down | and let it cool down. |
24