Invacare 3800, 3750 manual Opening/Closing the Tub Door, Tub Operation, Opening the Tub Door

Page 24

SECTION 3—OPERATION

Opening/Closing the Tub Door

WARNING

DO NOT open door when water is in the tub, serious injury could result.

NOTE: For this procedure, refer to FIGURE 3.1 on page 24.

Opening the Tub Door

1.Place one hand on top of the door handle.

2.Use the other hand to pull the safety pin out and release the door handle.

3.Lift up on the door handle and open the door.

Closing the Tub Door

WARNING

DO NOT fill or use tub if door does not lock - otherwise serious injury or dam- age may result. If the door does not lock, contact a qualified maintenance technician.

1.Close the door.

2.Push down on the door handle to lock door into position.

3.Lift the door handle to ensure the safety pin has engaged the door properly.

4.If the door handle lifts, repeat STEPS 1‐ 3.

Door Handle

Safety Pin

Door

FIGURE 3.1 Opening/Closing the Tub

Door

Tub Operation

WARNING

DO NOT stand in tub. Tub surfaces are slippery when wet. Serious injury may occur.

Before immersing the resident or operating the shower/massage wand, check that the water temperature is between 95° F and 105° F (35° C and 41° C) to avoid scalding the resident.

Disinfect the tub after each use to avoid resident contamination.

Recessed Side Entry Bath Tubs

24

Part No 1150695

Image 24
Contents Recessed Side Entry Bath Tubs Accessories Warning Table of Contents Troubleshooting Limited Warranty Special Notes These labels are located inside the service door Label LocationTypical Product Parameters 3750 3800Features 37503800 Important Safety Instructions InstallationGeneral Guidelines Electrical PlumbingMaintenance and Inspection DisinfectingOperation Parts Installation About InstallationParts Information Installing the Tub Stud Wall PocketShower Wand Plumbing Model 3750 Only Electrical and Plumbing Hook-up Locations on the Tub AB CElectrical Installation Electrical Installation Plumbing Installation Post Installation Checklist Plumbing InstallationOperation OverviewDisinfecting the Tub Using the Closed Loop System EquippedOpening/Closing the Tub Door Tub OperationOpening the Tub Door Closing the Tub DoorWater Temperature Setting Water Temperature SettingOperating Shower/Massage Wand Filling TubHydromassage Filling TubHydromassage Maintenance Safety Inspection ChecklistsMaintenance Overview Cleaning the Door Seal GasketCleaning Your Tub Removing Soap Film and Hard Water DepositsHeavy Duty Cleaning Procedure Required OZCalibrating the Disinfectant Flow Meter Calibrating the Disinfectant Flow Meter Maintenance Record Troubleshooting Symptoms Faults SolutionLimited Warranty USA