SECTION
Electrical
Be certain ALL electrical work is performed by a qualified electrician and is in compliance with local electrical codes.
ALL electrical connections are to be water resistant.
Grounding
DANGER
Grounding is required. The unit should be installed and grounded by a qualified service representative.
WARNING
The unit MUST be connected only to a supply circuit that is protected by a ground fault
A
Plumbing
Be certain ALL plumbing work is performed by a qualified plumber.
Some state and local plumbing codes require the installation of a “reduced pressure zone” (RPZ) assembly on both the incoming hot and cold supply lines to prevent back flow contamination into the potable water system. Invacare recommends you check local plumbing code requirements to determine if RPZs are required on your tub installation. RPZs are NOT supplied by Invacare and are to be provided by the customer if required.
The hot water supply to the tub MUST NOT exceed 110° F (43° C) to protect the user from scalding.
When operating water pressure exceeds 55 p.s.i., pressure regulators (similar to the ¾‐ inch Watts U5) MUST be installed.
Part No 1150695 | 11 | Recessed Side Entry Bath Tubs |