SECTION 4—MAINTENANCE

SECTION 4—MAINTENANCE

NOTE: Every six months or as necessary have a qualified technician perform a thorough inspection and servicing. Routine maintenance will extend the life and efficiency of your tub.

Safety Inspection Checklists

Initially

Ensure that the maximum water supply temperature does not exceed 115° F.

Ensure that the maximum water pressure does not exceed 65 PSI.

Verify the water flow temperature with a separate thermometer.

Visually inspect for cracks in the tub surface.

Ensure that the door handle locks the door closed.

Ensure that the safety pin properly engages the door.

Check that all labels are present and legible. Replace if necessary.

Before Each Use

Disinfect the tub.

Ensure that the maximum water supply temperature does not exceed 115° F.

Ensure that the maximum water pressure does not exceed 65 PSI.

Visually inspect for cracks in the tub surface.

Ensure that the door handle locks the door closed.

Ensure that the safety pin properly engages the door.

Every 80-150 Baths

Perform the heavy duty cleaning procedure on the tub. Refer to Heavy Duty Cleaning Procedure on page 31.

Every Six Months

Verify the water flow temperature with a separate thermometer.

Check that all labels are present and legible. Replace if necessary.

Clean RADA valve to remove lime deposits.

Part No 1150695

29

Recessed Side Entry Bath Tubs

Image 29
Contents Recessed Side Entry Bath Tubs Accessories Warning Table of Contents Troubleshooting Limited Warranty Special Notes Label Location These labels are located inside the service door3750 3800 Typical Product Parameters3750 Features3800 General Guidelines Important Safety InstructionsInstallation Plumbing ElectricalDisinfecting Maintenance and InspectionOperation Parts Information Parts InstallationAbout Installation Stud Wall Pocket Installing the TubShower Wand Plumbing Model 3750 Only AB C Electrical and Plumbing Hook-up Locations on the TubElectrical Installation Electrical Installation Plumbing Installation Plumbing Installation Post Installation ChecklistOverview OperationEquipped Disinfecting the Tub Using the Closed Loop SystemTub Operation Opening/Closing the Tub DoorOpening the Tub Door Closing the Tub DoorWater Temperature Setting Water Temperature SettingFilling Tub Operating Shower/Massage WandFilling Tub HydromassageHydromassage Safety Inspection Checklists MaintenanceCleaning the Door Seal Gasket Maintenance OverviewCleaning Your Tub Removing Soap Film and Hard Water DepositsRequired OZ Heavy Duty Cleaning ProcedureCalibrating the Disinfectant Flow Meter Calibrating the Disinfectant Flow Meter Maintenance Record Symptoms Faults Solution TroubleshootingUSA Limited Warranty