Chapter 2: Installing the monitor

To connect the AC power cable:

1.If the monitor has a hatch, make sure that the hatch is open (it should have been opened when connecting the video cable). The hatch must be open to access connectors.

2.Connect the power cable to the socket on the side of the monitor.

3.Connect the other end of the cable to a standard 15-ampere (or less) branch circuit.

AC power supply with fixed field wiring(ET 350R monitor and ET 350L3 monitor only)

Warning

To reduce the risk of fire or electric shock which could result in serious personal injury or death, wiring installation should be done by a licensed journeyman electrician and must comply with federal and local electrical codes.

To reduce the risk of electric shock which could result in serious personal injury or death, provide adequate strain relief for all communication and power cables.

To connect the power supply with fixed field wiring, follow these steps:

1.Make sure the AC power supply is disconnected at its source.

2.Strip about 1/8-inch (3 mm) of insulation from the end of the neutral, ground, and line wires.

Warning

To reduce the risk of fire or electric shock which could result in serious personal injury or death, when connecting power with fixed field wiring for the ET 350L3 monitor and the ET 350R monitor, the power cable must be double insulated. A clear, flexible insulator (supplied) must cover the portion of the cable that is not double insulated when connecting to a terminal strip connector.

3.Measure and, if necessary, cut the clear, flexible insulator (supplied) so that it will cover the unjacketed portion of the power cable and overlap the jacketed portion by at least one inch.

4.Insert the stripped end of the AC power cable wire (neutral) into the terminal strip’s side connector (Figure 6).

17