Related messages (continued)

Troubleshooting

Related messages (continued)

Message

CHECK THE AIR FLOW

CLEAN THE AIR FILTER

POWER OFF FIRST,

THEN CLEAN THE AIR

FILTER.

AFTER CLEANING THE AIR

FILTER,

RESET THE FILTER TIMER.

Description

The internal portion temperature is rising.

Please turn the power OFF, and allow the unit to cool down at least 20 minutes. After having confi rmed the following items, please turn the power ON again.

Is there blockage of the air passage aperture?

Is the air fi lter dirty?

Does the peripheral temperature exceed 35°C?

If the same indication is displayed after the treatment, please set the HIGH at the FAN SPEED of the item SERVICE of the OPTION menu.

A note of precaution when cleaning the air filter.

Please immediately turn the power OFF, and clean or change the air fi lter by referring to the “Air Filter” section of this manual. After you have cleaned or changed the air fi lter, please be sure to reset the fi lter timer ( 39)

NOTE • A lamp has a fi nite product life. Lamps are characterized by the fact that, after long hours of usage, a lamp will no longer light up, or the lamp will break or burst, etc. This projector is equipped with an automatic shut-down function, such that the power will automatically be turned off when lamp usage time has reached 2000 hours. Please be aware, however, that among lamp types, there are major differences in product lifetimes; a lamp may thus fail to light even prior to the functioning of the automatic shut-down function of this projector.