CALIBRATION

CALIBRATION

When to calibrate

Calibration may be required when the scale initially installed, if it is moved a substantial distance, or in accordance with local regulations. This is necessary because the weight of a mass in one location is not necessarily the same in another location. Also, with time and use, mechanical deviations can occur.

The position of the [UNITS] key for the Calibration

The “g” model has no

Location of the [UNITS] key

[UNITS] key, but the key

used to enter calibration is

 

located under the overlay.

 

Calibration using a calibration weight

Calibrate the scale with calibration weights at full capacity.

HL-1000WP :

1000g

± 0.1g

HL-300WP:

300g

± 0.02g

HL-3000WP :

3000g

± 0.2g

HL-3000LWP :

3000g

± 0.2g

1. Enter calibration mode.

Press the [ON/OFF] key to turn the power off.

While pressing both the [RE-ZERO] key and the [UNITS] key, press the [ON/OFF] key. Cal will be displayed.

2. Zero calibration

Press the [RE-ZERO] key to calibrate zero. Cal 0 will be displayed. Wait for the stable indicator to be displayed and press the [RE-ZERO] key. Cal f will be displayed in a few seconds. If only the zero calibration is to be done, press the [UNITS] key. The scale will display end and automatically return to the weighing mode.

3. Span calibration

Place the weight at the center of the platform for accurate calibration. Wait for the stable indicator to be displayed and then press the [RE-ZERO] key. The display will show end and the scale will automatically return to the weighing mode.