CALIBRATION
When to calibrate
Calibration may be required when the scale initially installed, if it is moved a substantial distance, or in accordance with local regulations. This is necessary because the weight of a mass in one location is not necessarily the same in another location. Also, with time and use, mechanical deviations can occur.
The position of the [UNITS] key for the Calibration
The “g” model has no | Location of the [UNITS] key |
[UNITS] key, but the key | |
used to enter calibration is |
|
located under the overlay. |
|
Calibration using a calibration weight
Calibrate the scale with calibration weights at full capacity.
1000g | ± 0.1g | |
300g | ± 0.02g | |
3000g | ± 0.2g | |
3000g | ± 0.2g |
1. Enter calibration mode.
Press the [ON/OFF] key to turn the power off.
While pressing both the
2. Zero calibration
Press the
3. Span calibration
Place the weight at the center of the platform for accurate calibration. Wait for the stable indicator to be displayed and then press the
5