8.2AUTO CALIBRATION FOR ALL OTHER MODELS
•Before entering calibration mode, press [On/Off] to turn the power on for 1 minute.
Enter calibration mode
•Press [On/Off] to turn the power off.
•Press [Unit] first and then press [On/Off], while keeping the [Unit] key pressed. The display will show the last calibration value used (For example, “ L – 1000”), press [Tare] to choose between the two calibration weight values. When you have made your selection press [Unit] and the display will show the A/D value.
Calibration
•Wait for the stable indicator to be displayed,
•Press [Unit]
•Place the required weight on the platform. This is the weight that you selected as your calibration weight value earlier.
•After the stable indicator is displayed press [Unit]
•The display will show "00SAVE". The calibration in complete now.
•Press [On/Off] to turn the power off. Press [On/Off] to turn on the power again,
•Place full weight (eg.2000g for AQT2000) on the platform to verify weighing is correct. If not, repeat the calibration procedure.
9.0ERROR MESSAGES
EEEEHI | Overload | ||
|
|
|
|
EEEELO | A/D value too low | ||
|
|
|
|
|
|
| Low battery |
|
|
| |
|
|
|
|
Note: Calibration may be required before weighing. Read section 8.0 on "CALIBRATION" first and if necessary calibrate your balance for accurate weighing.